![Agilent Technologies Agilent 3499A Скачать руководство пользователя страница 345](http://html.mh-extra.com/html/agilent-technologies/agilent-3499a/agilent-3499a_user-manual_2867269345.webp)
345
Chapter 9 Specifications
44477A Form-C Relay Module
4
9
44477A Form-C Relay Module
INPUT CHARACTERISTICS
Total Channels:
7
Maximum Voltage:
Terminal-Terminal or
Terminal-Chassis:
250 V dc or ac rms,
350 V ac peak
Maximum Current:
Per Channel:
Per Module:
2 A, dc or ac rms
14 A, dc or ac rms
Maximum Power:
Per Channel:
Per Module:
60 W dc; 500 VA ac
420 W dc; 3500 VA ac
Maximum Overvoltage
Transients:
1400 V
pk
Thermal Offset:
< 3
µ
V per channel
Initial Closed Channel
Resistance:
< 1
Ω
Relay Life:
Dry Load of < 300 mA & < 10 V:
Maximum Rated Load:
10
8
10
5
Maximum Scan Rate:
a
a.
Using the 44474A external increment & channel closed, display off.
43 Chans/sec
DC ISOLATION
Open Channel,
Channel-Channel:
(with 1 channel closed)
≤
(40
o
C, 60% RH):
≤
(40
o
C, 95% RH):
> 10
11
Ω
>10
9
Ω
Channel-Chassis:
(with 1 channel closed)
≤
(40
o
C, 60% RH):
≤
(40
o
C, 95% RH):
> 5 x 10
11
Ω
> 10
10
Ω
Содержание Agilent 3499A
Страница 13: ...1 1 Quick Start ...
Страница 27: ...2 2 Front Panel Operation ...
Страница 55: ...3 3 System Overview ...
Страница 77: ...4 4 Features and Functions ...
Страница 113: ...5 5 Remote Interface Reference ...
Страница 164: ...164 5 ...
Страница 165: ...6 6 Error Messages ...
Страница 173: ...7 7 Plug in Modules ...
Страница 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Страница 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Страница 268: ...268 Chapter 7 Plug in Modules 44478A B 1 3 GHz Dual 4 to 1 MUX Modules 7 The 44478A B is shown below ...
Страница 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Страница 288: ...288 7 ...
Страница 289: ...8 8 Application Programs ...
Страница 299: ...9 9 Specifications ...
Страница 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...