To Output an FSK Waveform
4
2
To Output an FSK Waveform
You can configure the waveform generator to “shift” its output frequency
between two preset values using FSK modulation. The rate at which the
shift happens is the FSK rate, and this FSK rate is determined by the
internal rate generator or the signal level on the rear-panel
Ext Trig
connector. The two frequencies that the signal shifts between are called
the “carrier frequency” and the “hop frequency.”
For this example,
you will set the “carrier” frequency to 3 kHz and the “hop” frequency to
500 Hz, with an FSK rate of 100 Hz.
1 Select the function, frequency, and amplitude of the carrier.
Press
and then press the
Sine
softkey. Press the
Frequency
,
Amplitude
, and
Offset
softkeys to configure the carrier waveform.
For
this example, select a 3 kHz sine wave with an amplitude of 5 Vpp, with 0
offset.
2 Select FSK.
Press
and then select
FSK
using the
Type
softkey. Then press the
Modulate
softkey to turn modulation On. Notice the status message
“FSK Modulated” at the top left of the display.
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...