LAN Configuration
SYSTem:COMMunicate:LAN:GATeway "<
address
>"
SYSTem:COMMunicate:LAN:GATeway? [{CURRent|STATic}]
Assigns a default gateway for the instrument. The specified IP Address sets the default gateway which allows the
instrument to communicate with systems that are not on the local subnet. Thus, this is the default gateway where
packets are sent which are destined for a device not on the local subnet, as determined by the Subnet Mask setting. If
DHCP is enabled (SYSTem:COMMunicate:LAN:DHCP ), the specified default gateway is not used. However, if the DHCP
server fails to assign a valid IP address, the currently configured default gateway will be used. Contact your LAN admin-
istrator for details.
If you change this setting, you must send
to activate the new
setting.
Parameter
Typical Return
Command: "
", default "0.0.0.0"
Query: {CURRent|STATic}, default CURRent
"198.105.232.1"
Set default gateway address:
SYST:COMM:LAN:GATEWAY "198.105.232.1"
SYST:COMM:LAN:UPD
l
CURRent: read address currently being used by the instrument.
l
STATic: read static address from non-volatile memory. This address is used if DHCP is disabled or unavailable.
l
Set to "0.0.0.0" when instrument is shipped from factory or after SYSTem:SECurity:IMMediate.
SYSTem:COMMunicate:LAN:HOSTname "<
name
>"
SYSTem:COMMunicate:LAN:HOSTname? [{CURRent | STATic}]
Assigns a hostname to the instrument. A hostname is the host portion of the domain name, which is translated into
an IP address. If Dynamic Domain Name System (DNS) is available on your network and your instrument uses DHCP,
the hostname is registered with the Dynamic DNS service at power-on. If DHCP is enabled (SYS-
Tem:COMMunicate:LAN:DHCP ), the DHCP server can change the specified hostname.
If you change this setting, you must send
to
activate the new setting.
Parameter
Typical Return
String of up to 15 characters.
Must start with letter (A-Z)
May contain letters, numbers (0-9), or dashes ("-")
"LAB1-33522A"
Define a hostname:
SYST:COMM:LAN:HOST
l
Set to "A-33521A-nnnnn" or "A-33522A-nnnnn", where nnnnn is the last five digits of the instrument's serial
number, when the instrument is shipped from the factory or after
310
Agilent 33500 Series Operating and Service Guide
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......