18
Technical Information
Air Cooling Kit
292/306
Turbo-V 903 Pumping System User Manual
/ 87-901-016-01
(A)
Air Cooling Kit
TV 903 Pumping System
An air cooling kit (mod. 5218 NM) is available to improve the TV 903 Pumping
System cooling during heavy operational conditions.
Single fan specifications:
air flow: 187 m
3
/h
input voltage: 48 Vdc
dimensions: 127 x 127 x 38 mm (5 x 5 x 1.5 in.)
power: 5 W
To supply the fan it is necessary to connect the fan supply to the P4 connector of
the controller with the special shielded cable included into the kit to reduce noises
and to accomplish to EMC norms.
Figure 21
Pumping system
Содержание 969-6500
Страница 4: ...Turbo V 903 Pumping System 4 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 14: ...Contents 14 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 58: ...3 Mode d emploi Mise au rebut 58 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 88: ...5 Manual de Istruções Eliminação 88 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 104: ...6 Bedrijfshandleiding Afvalverwerking 104 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 118: ...7 Istruktionsbog Bortskaffelse 118 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 176: ...11 Oδηγιεσ Χρησεωσ Χώνευση 176 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 218: ...14 Návod k Použití Likvidace 218 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 232: ...15 Návod na Obsluhu Likvidácia 232 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 258: ...17 Instructions for Use Disposal 258 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 300: ...18 Technical Information Accessories and Spare Parts 300 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......