![Agilent Technologies 9499225 Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/agilent-technologies/9499225/9499225_user-manual_2868747090.webp)
Technical Information
RS 232/RS 485 Communication Description
5
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (
H.00
)
90/104
Type
Length
Value
Description
Out of Range
1 byte
(0x34)
the value expressed during a write
command is out of the range value of the
specified window
Win Disabled
1 byte
(0x35)
the specified window is Read Only or
temporarily disabled (for example you
can’t write the Soft Start when the Pump
is running)
NOTE
Using the RS 485 interface, the message structure remains identical to the one
used for the RS 232 interface, the only difference being that the value assigned
to the ADDRESS <ADDR>.
Examples
Command: START
Source: PC
Destination: Pump
02
80
30 30 30 31
31 03
42 33
STX ADDR
WINDOW
WR ON ETX
CRC
Source: Pump
Destination: PC
02
80
06
03
38 35
STX ADDR ACK ETX CRC
Command: STOP
Source: PC
Destination: Pump
02
80
30 30 30 31
30
03
42 32
STX ADDR WINDOW
WR OFF ETX CRC
Содержание 9499225
Страница 8: ...Contents MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 8 104...
Страница 33: ...Gebrauchsanleitung Entsorgung 2 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 33 104...
Страница 46: ...Mode d emploi Mise au rebut 3 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 46 104...
Страница 59: ...Instructions for Use Disposal 4 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 59 104...