Instructions for Use
Installation
4
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (
H.00
)
52/104
Installation
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
temperature: from +12 °C to +40 °C
relative humidity: 0 – 95 % (non-condensing).
CAUTION!
Before starting the pump, fill up with lubricating oil as the pump is delivered
empty.
NOTE
It is important that the oil level stays within the range MIN and MAX that is
indicated by the level gauge on the side of the pump.
CAUTION!
It is mandatory to leave ample free space all around the pump to allow for
correct air circulation; it is forbidden to insert the pump into a closed space
that restricts air circulation. The internal security system may notify of an
error condition when the environmental operating temperature of the
pump exceeds 40 °C.
WARNING!
Keep the caps in place on the suction and exhaust flanges and do not switch
the pump on until the flanges are appropriately connected respectively to the
exhaust instrument and to the exhaust line.
Содержание 9499225
Страница 8: ...Contents MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 8 104...
Страница 33: ...Gebrauchsanleitung Entsorgung 2 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 33 104...
Страница 46: ...Mode d emploi Mise au rebut 3 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 46 104...
Страница 59: ...Instructions for Use Disposal 4 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 59 104...