FETCh:PVTime
FETCh:PVTime:POWer:TIME[:OFFSet]:SDEViation?
FETCh:PVTime:ICOunt?
FETCh:PVTime:INTegrity?
Function
Queries the standard deviation, from a number of multi-measurements, at each user-settable
time offset appended to this command. Specified time values must correspond to user-settable
time offsets that are currently turned on, and must be rounded to the same values. (9.91 E+37
(NAN) will be returned for specified offsets that do not correspond to offsets currently turned on).
Power levels are relative to the power versus time carrier power measurement.
The
command sets up the number of offsets that are turned on
and their time values. Measurements will be returned by this query in the same order they are
listed in the command.
Query
Power levels:
• Range: 0 dBc to +100 dBc and 9.91 E+37 (NAN)
• Resolution: 0.001 dB
Time offsets:
• Range: Up to 12 time offset values, corresponding to entries in the Power vs Time table of
user-defined time offsets currently turned on. The default units are s (seconds).
• Resolution: Rounded to the same value as displayed in the Power vs Time table and returned
by the
query.
Programming Example
OUTPUT 714;"FETCH:PVTIME:POWER:TIME:OFFSET:SDEVIATION? 0 US, 570.8 US"
!Returns the standard deviation of power versus time measurements
at the 0.0 ms and 570.8 ms offsets.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Queries the intermediate count of power versus time multi-measurements completed.
Query
Range: 0 to 999
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Queries the integrity indicator for the power versus time measurement. Zero indicates a normal
result.
See
“Integrity Indicator” on page 125
for descriptions of non-zero integrity indicators.
Query
Range: 0 to 16
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Содержание 8960 Series 10
Страница 26: ...26 Contents ...
Страница 363: ...363 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual Chapters hpib_read fm READ ...
Страница 500: ...500 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual ch_gen_usage fm General Usage ...
Страница 574: ...574 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual ch_error_messages fm Error Messages ...