FETCh:PVTime
FETCh:PVTime[:ALL]?
Function
Queries integrity indicator, mask pass/fail indicator, power versus time (PvT) transmit power
(average), and PvT power (maximum) at up to 12 time offsets.
The number of PvT measurement offsets that will be returned can be queried using the
“SETup:PVTime:TIME:POINts?” on page 430
. The time offsets are set up using the command
“SETup:PVTime:TIME[:OFFSet]” on page 429
Query
Integrity indicator
• Range: 0 to 16
• Resolution: 1
Power versus time mask pass/fail
• Range: 0 (pass) or 1 (fail) and 9.91 E+37 (NAN)
Power versus time TX carrier power
• Range:
−
100 dBm to 100 dBm and 9.91 E+37 (NAN)
• Resolution: 0.01 dB
Power versus time power (maximum)
• Range:
−
100 dBm to 100 dBm and 9.91 E+37 (NAN)
• Resolution: 0.01 dB