Maintaining the Purged Packed Inlet
To Clean the Purged Packed Inlet
90
Maintaining Your GC
To Clean the Purged Packed Inlet
1
Gather the following:
•
Replacement O-ring, see
“Consumables and Parts for the Purged Packed Inlet”
•
Replacement glass liner
•
Replacement septum
•
Solvent that will clean the type of deposits in your inlet
•
Compressed, filtered, dry air or nitrogen
•
Beaker
•
Cleaning brushes—The FID cleaning kit (part number 9301-0985) contains appropriate
brushes
•
Lint-free gloves
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands
.
3
Remove the column.
4
Remove the septum nut and septum.
5
Remove the top insert assembly.
6
Remove the glass liner and O-ring.
7
If used, remove the adapter.
8
Ultrasonically clean the septum nut, top insert assembly, and adapter (if used) in a suitable
solvent.
9
Place a beaker in the oven under the inlet to catch the solvent.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
10
Soak a cleaning brush with the solvent and vigorously scrub the interior walls of the inlet.
11
Blow the inside of the inlet dry with compressed air or nitrogen.
12
Install the adapter, if used. See
“To Install an Adapter on the Purged Packed Inlet”
13
Install the glass liner and O-ring. See
“To Change the Glass Liner on the Purged Packed
14
Install the top insert assembly and finger-tighten.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...