Maintaining the GC
To Change the Hydrogen Sensor Calibration Gas Cylinder
Maintaining Your GC
27
To Change the Hydrogen Sensor Calibration Gas
Cylinder
Replace the calibration gas cylinder when you cannot adjust it to provide the correct flow rate
during calibration, or if it expires.
WARNING
Connect only the Agilent calibration gas to the hydrogen sensor fitting. The calibration gas is
2% hydrogen in air and is not potentially explosive. Gases with higher concentrations of
hydrogen can create a dangerous condition in the oven or damage the hydrogen sensor.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. While hydrogen
sensor calibration can be performed at any temperature, cool the oven and detector to safe
handling temperatures, < 40 °C, before continuing.
1
Make sure the old cylinder is empty.
2
Turn the pressure regulator completely off and set the output pressure as low as possible
(full CCW).
3
Remove the pressure regulator from the old cylinder and install it onto the new cylinder.
4
Install the new cylinder into the stand.
5
Turn on the pressure in the new cylinder.
6
Open the GC oven door and connect a flowmeter tube to the sensor tube in the oven.
7
Select
Settings > Calibration > Hydrogen Sensor > Start Calibration Cycle? > On/Yes
. The
calibration cycle begins. The hydrogen sensor module will wait to stabilize, then will begin
to send calibration gas across the sensor.
8
While continuing to measure the flow rate from the tube, adjust the pressure regulator on
the calibration gas cylinder until the flow rate is approximately 30 mL/min. Remove the
flowmeter and close the oven door.
9
Allow the calibration cycle to complete (approximately 5 minutes total).
10
Check for leaks.
Connect the flow meter
to the hydrogen sensor
tube next to the oven
heater shroud
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...