12 Maintaining the TCD
To Bakeout Contaminants from the TCD
Maintaining Your GC
207
To Bakeout Contaminants from the TCD
The bakeout can be performed with the column installed or the detector capped.
CAUTION
If the column is not installed, you must turn off the TCD filament and cap the detector column
fitting to prevent irreparable damage to the filament caused by oxygen entering the detector.
1
If the column is not installed, cap the detector.
WARNING
Be careful! The detector may be hot enough to cause burns. If the detector is hot, wear
heat-resistant gloves to protect your hands.
2
Turn off the TCD filament.
3
If the column is attached to the inlet, maintain inert carrier gas flow through the column.
WARNING
If using hydrogen as a carrier gas, turn off the hydrogen supply and cap the end of the column
to prevent an oven explosion.
4
Set the reference gas flow rate between 20 and 30 mL/min.
5
Set the detector temperature to 375 °C.
6
Hold at 375 °C for several hours.
7
If the column is uninstalled, install it. See
“To Install a Capillary Column in the TCD”
8
Load the analytical method.
WARNING
Be careful! The oven or detector fittings may be hot enough to cause burns.
9
Allow the oven, inlet, and detector to equilibrate at operating temperature, then re-tighten
the fittings.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...