![Agilent Technologies 8890 Скачать руководство пользователя страница 141](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678141.webp)
8
Maintaining the MMI
To Bakeout Contaminants from the MMI
Maintaining Your GC
141
To Bakeout Contaminants from the MMI
1
Put the inlet into split mode.
2
Set the column flow to the normal operating value, or set the capillary column gas velocity
to 30 cm/s.
3
Set the inlet split vent flow to 200 mL/min.
4
Purge the column with carrier flow for at least 10 minutes before heating the oven.
5
If the column is attached to the detector, set the detector 25 °C above normal operating
temperature.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If they are hot,
wear heat-resistant gloves to protect your hands.
6
If the column is not attached to the detector, cap the detector fitting.
7
Set the inlet temperature to 300 °C or 25 °C above the normal operating temperature to
bakeout contaminants from the inlet, mostly through the split vent.
8
Set the column oven 25 °C above the GC method final oven temperature to bake
contaminants from the column. Do not exceed the column manufacturer’s maximum
temperature limit.
9
Bakeout for 30 minutes or until the detector baseline is free of contamination peaks.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...