Maintaining the MMI
To Clean the Multimode Inlet
Maintaining Your GC
139
To Clean the Multimode Inlet
This procedure explains how to clean the Multi Mode Inlet (MMI). Depending on the inlet mode
used, the liner installed, and the cleanliness of the sample, the frequency of cleaning may
range from weekly to monthly. Start with a visual inspection of the inlet bottom whenever
changing the liner. A small ring of material will collect at the bottom of the inlet when injecting
dirty samples such as food extracts or solid waste. Try an initial cleaning schedule of every two
weeks for dirty samples and every two months for clean samples and adjust as needed.
1
Gather the following:
•
Cleaning kit, multimode inlet (G3510-60820). Contains 5 each abrasive swabs and
5 each cotton swabs.
•
Solvent for cleaning (at least 25 mL of acetone, isopropanol, or hexane, whichever is
most appropriate for your sample matrix).
•
100 mL beaker or similar container to the catch solvent rinse.
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands
.
3
Turn off the GC.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
“To Change the Liner and O-Ring on the MMI”
5
Remove the column from the inlet. See
“To Install a Capillary Column with the Multimode
6
Place the glass beaker or other container in the GC oven, below the MMI inlet, so it will
catch any solvent that drips through the inlet body.
7
Moisten a cotton swab with solvent and wipe the inside bottom seal of the inlet repeatedly
to remove any loose material from the MMI. If desired, the cotton swab can be used to
clean the bore of the inlet. For most samples, this is not necessary. Dispose of the swab
after a single use.
8
Using a dry abrasive swab, place the swab all the way to the bottom of the inlet and with
moderate pressure, twist the swab forwards and backwards while it is in contact with the
bottom of the MMI (approximately 20 cycles should be sufficient). The abrasive swab can
be used multiple times.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...