![Agilent Technologies 8890 Скачать руководство пользователя страница 100](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678100.webp)
6
Maintaining the Purged Packed Inlet
To Install Ferrules on a Packed Metal Column
100
Maintaining Your GC
To Install Ferrules on a Packed Metal Column
1
Gather the following:
•
Wrenches
•
Stainless steel male Swagelok fitting, 1/4- or 1/8-inch od
•
Brass Swagelok nut and ferrule set, see
“Consumables and Parts for the Purged
•
Lint-free gloves
2
Verify that the column end is cut square and is free of burns and deformation.
3
Secure the fitting in a bench vise.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
Assemble a Swagelok nut and ferrules onto the column.
5
Fully insert the column into the vise-held fitting, then withdraw 1–2 mm. Finger-tighten the
nut.
6
Tighten the nut an additional 3/4 turn with a wrench (for a 1/8-inch column) or an
additional 1-1/4 turn (for a 1/4-inch column).
7
Unscrew the column nut from the vise-held fitting and remove the column. Ferrules should
now be set in place on the column with the column end correctly positioned.
Front ferrule
Back ferrule
Column nut
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...