![Agilent Technologies 8712ES Скачать руководство пользователя страница 92](http://html.mh-extra.com/html/agilent-technologies/8712es/8712es_programmers-manual_2867956092.webp)
6-22
Programmer’s Guide
Trace Data Transfers
Internal Measurement Arrays
CH[1|2]BFWD
and
CH[1|2]RFWD
. Raw data for AUX INPUT is not
available via GPIB. Use the corrected data array to access AUX INPUT
data.
Table 6-4
Raw Data Arrays
Ratio Calculations
These are performed if the selected measurement is a ratio (e.g. A/R or
B/R). This is simply a complex divide operation. If the selected
measurement is absolute (e.g. A or B), no operation is performed.
Error Correction
Error correction is performed next if correction is turned on. Error
correction removes repeatable systematic errors (stored in the error
coefficient arrays) from the raw arrays. The operations performed
depend on the selected measurement type.
Selected Measurement
Raw Arrays
Transmission (B/R)
B =
CH[1|2]BFWD
, R=
CH[1|2]RFWD
Reflection (A/R)
A =
CH[1|2]AFWD
, R=
CH[1|2]RFWD
A
A =
CH[1|2]AFWD
B
B =
CH[1|2]BFWD
R
R =
CH[1|2]RFWD
Power (B*)
B*=
CH[1|2]BFWD
Conversion Loss (B*/R*)
B*=
CH[1|2]BFWD
, R*=
CH[1|2]RFWD
R*
R*=
CH[1|2]RFWD
AM Delay (Y/X)
Y =
CH[1|2]BFWD
, X =
CH[1|2]RFWD
X
X
= CH[1|2]RFWD
Y
Y =
CH[1|2]BFWD
Y/R*
Y
= CH[1|2]BFWD
, R* =
CH[1|2]RFWD
Y/X, X/Y
Y =
CH[1|2]BFWD
, X =
CH[1|2]RFWD
Содержание 8712ES
Страница 11: ...1 1 1 Introduction to GPIB Programming ...
Страница 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Страница 36: ...3 1 3 Passing Control ...
Страница 39: ...4 1 4 Data Types and Encoding ...
Страница 46: ...5 1 5 Using Status Registers ...
Страница 71: ...6 1 6 Trace Data Transfers ...
Страница 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Страница 99: ...7 1 7 Using Graphics ...
Страница 105: ...8 1 8 Front Panel Keycodes ...
Страница 111: ...9 1 9 Introduction to SCPI ...
Страница 129: ...10 1 10 Menu Map with SCPI Commands ...
Страница 268: ...12 1 12 SCPI Conformance Information ...
Страница 290: ...13 1 13 SCPI Error Messages ...