Programmer’s Guide
3-3
Passing Control
Passing Control
The following is a procedure for passing control:
1. Send the controller's GPIB address to the analyzer with the
*PCB
command.
2. Clear the analyzer's status registers with the
*CLS
command.
3. Enable the analyzer's status registers to generate a service request
when the Operation Complete bit is set. (Send
*ESE
with a value of 1
and
*SRE
with a value of 32.)
4. Enable the controller to respond to the service request.
5. Send the command that requires control of the bus followed by the
*OPC
command.
6. Pass control to the analyzer and wait for the service request. The
service request indicates that the command has been completed and
control has been passed back to the controller.
NOTE
For this procedure to work properly, only the command that requires
control of the bus should be pending. Other overlapped commands should
not. For more information on overlapped commands, see
Содержание 8712ES
Страница 11: ...1 1 1 Introduction to GPIB Programming ...
Страница 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Страница 36: ...3 1 3 Passing Control ...
Страница 39: ...4 1 4 Data Types and Encoding ...
Страница 46: ...5 1 5 Using Status Registers ...
Страница 71: ...6 1 6 Trace Data Transfers ...
Страница 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Страница 99: ...7 1 7 Using Graphics ...
Страница 105: ...8 1 8 Front Panel Keycodes ...
Страница 111: ...9 1 9 Introduction to SCPI ...
Страница 129: ...10 1 10 Menu Map with SCPI Commands ...
Страница 268: ...12 1 12 SCPI Conformance Information ...
Страница 290: ...13 1 13 SCPI Error Messages ...