17 Technical Information
Interconnections – Controller Model 9698978M004
250/294
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
Tab. 3
Signal characteristics and use
PIN N.
SIGNAL NAME
IN-/OUTPUT
1
START/STOP (+)
IN
2
START/STOP (-)
IN
3
INTERLOCK (+)
IN
4
INTERLOCK (-)
IN
5
SPEED SETTING (+)
IN
6
SPEED SETTING (-)
IN
7
SOFT START (+)
IN
8
SOFT START (-)
IN
9
+24 Vdc
OUT
10
SPARE
11
PROGRAMMABLE SET POINT
OUT
12
SPARE
13
FAULT OUTPUT
OUT
14
PROGRAMMABLE ANALOG SIGNAL (+)
OUT
15
GROUND
PROGRAMMABLE ANALOG SIGNAL (-)
OUT
Signal Description
Start/Stop
: input signal to start or stop the pump. With the supplied
cover connector the START/STOP (+) signal is connected to the +24
Vdc pin and the START/STOP (-) signal to the GROUND pin: in this
condition the pump automatically starts as soon as the controller
recognises the input supply ("Plug & Pump").
Interlock
: input signal to control the pump rotation. With the
supplied cover connector the interlock (+) signal is connected to the
+24 Vdc pin and the interlock (-) signal to the GROUND pin.
Speed setting
: PWM input signal to set the pump speed. The PWM
signal characteristics must be the following:
frequency: 100 Hz +/-20%
Содержание 8698961R001
Страница 3: ...Turbo V 1K G Retrofit Turbo V 1K G Retrofit User Manual 87 900 996 01 B 3 294 Turbo V 1K G Retrofit...
Страница 4: ...Turbo V 1K G Retrofit 4 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 14: ...Contents 14 294 Turbo V 1K G Retrofit User Manual 87 900 996 01 B...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......