2-29
Using the Instrument
Analyzing Measurement Data
Using Trace Math to Measure Wavelength Drift
1
From the front panel, press
Traces
>
Active Trace
>
TrA
.
2
Press
Single Sweep
,
Bandwidth Sweep
,
Single Sweep
to update Trace A
then press
Traces
,
Update A off
.
3
Press
Active Trace
>
TrB
.
4
Press
Sweep
>
Repeat Sweep On
to continuously update the measured
response on Trace B.
5
Press
Traces
>
Trace Math
,
Default Math Trace C
>
Log Math C = A – B
.
You can now monitor the wavelength drift of your device over time.
Refer to “Log Math C=A–B” on page 3-35
, and