Using the Instrument
Setting Up Measurements
Triggering a Measurement
Triggering a measurement synchronizes the start of the sweep to an internally
generated trigger signal. Internal triggering ensures continuously triggered
sweeps with the shortest delay between sweeps.
In some measurements, the spectrum at a particular time within the modula-
tion period is more important than the average spectrum. Gated triggering can
be used to synchronize the data acquisition portion of the OSA to a gating trig-
ger connected to the rear-panel
EXT TRIG IN
connector. Gated triggering
requires a TTL-compatible signal with a minimum of 0 Vdc and a maximum of
+5 V.
Gated triggering ignores the spectrum when the trigger input is low. It usually
is used in conjunction with the Max Hold function during several sweeps.
Gated triggering is used to select data samples containing valid information.
When the gating signal is high, the data sample is accepted. When the gating
signal is low, the data sample is replaced by a data point with a value of
–200 dBm. Processing continues according to the functions selected, such as,
video bandwidth, max hold, and so forth.
If the low level exists for longer that the time needed for the grating to move
from one trace point to the next, then the trace will have “gaps”. There are
two ways to eliminate the gaps. You can increase the sweep time to at least:
(1.2–2 times the product trace length)
×
(the longest “low level” period)
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...