3-40
Function Reference
Marker Setup
Setup Panel
Selections
Normal Marker Units
Sets the X-axis immediately displayed marker information for frequency or
wavelength when the instrument is in a non-zero span. This setting controls
only the normal marker X-axis and the delta reference immediately displayed
information. This setting controls all four normal markers. The bandwidth and
delta offset markers have their own settings.
BW Marker Units
Sets the bandwidth marker X-axis immediately displayed information for fre-
quency or wavelength when the instrument is in a non-zero span. The delta
markers have their own setting. This setting controls all four bandwidth mark-
ers.
Delta Marker Units
Sets the delta marker X-axis immediately displayed information for frequency
or wavelength when the instrument is in a non-zero span. The bandwidth
markers have their own setting. This setting controls all four delta markers.
Normal/Delta Marker Interpolation
Turns the normal/delta marker interpolation on or off. When interpolation is
ON
, the normal/delta markers will be placed at the exact wavelength or fre-
quency value selected if it is within the trace range. The marker will linearly
interpolate between two trace data points. The default state is off.
This setting controls the interpolation state for all four markers when in the
normal or delta mode. When interpolation is off, the marker will be placed at
the trace point nearest the requested value.
Bandwidth Marker Interpolation
Turns the bandwidth marker interpolation on or off. When interpolation is
ON
,
the bandwidth markers will be placed at the exact number of dB (NDB) from
the normal marker if within the trace range. The position of the marker will be
linearly interpolated between two trace data points. The default state is
ON
. If
interpolate is
OFF
, for negative NDB values, the bandwidth markers will be at
values closest to and more negative than the NDB value. For positive NDB val-
ues, the bandwidth markers will be at values closest to and more positive than
the NDB values. This will typically result in a wider bandwidth measurement.
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...