3-24
Function Reference
Display Setup
Title
Turns the title
ON
or
OFF
on the display.
Active Function Area Assist
Automatically sets the Active Function Area (when function is turned on) to
the first or top softkey. For example, when you press the front-panel
Band-
width/Sweep
key, the Active Function Area Assist function will set the Active
Function Area to resolution bandwidth. This means you can immediately
enter the desired resolution bandwidth. There is no need to press the
Res BW
softkey
Path Change Instructions Panels
For the Agilent 86144B/86146B Filter Mode only
Allows the choice of disabling the external path align setup instructions.
Filter Mode instruction Panels
For the Agilent 86144B/86146B Filter Mode only
Allows the choice of disabling the auto align, auto measure, and take sweep
setup instructions.
Key Path
System > More System Functions > Display Setup
Remote
Commands
DISPlay:WINDow:TEXT:DATA <string>
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...