Addressing and Initializing the Instrument
4
-
3
Addressing and Initializing the Instrument
The GPIB address of the Agilent 86121A is configured
at the factory to a value of 20. You must set the output
and input functions of your programming language to
send the commands to this address. You can change the
GPIB address from the front panel as described in
change the GPIB address” on page 4-4
.
Remote mode and front-panel lockout
Whenever the instrument is controlled by a computer,
the
Remote
message is displayed on the screen of the
instrument and the softkey menu is blanked except for
the
LOCAL
softkey. This softkey can be pressed by the
user to restore front panel control of the instrument.
You can specify a local lockout mode by remote com-
mand that prevents the
LOCAL
softkey from being dis-
played. If the instrument is in local lockout mode, all the
softkeys may be blanked. For example, if the instru-
ment is first placed in local lockout mode and then
placed in remote mode, no softkeys are displayed.
Consult the documentation for your programming envi-
ronment to determine which commands are used to put
an instrument in the remote and local lockout modes.
These are not Agilent 86121A commands; they control
GPIB control lines and do not send any characters to
the 86121A.
Set single acquisition mode
Whenever the instrument is under remote program-
ming control, it should be in the single measurement
acquisition mode. This is automatically accomplished
when the *RST common command is used. The *RST
command initializes the instrument to a preset state:
Pressing the green
Preset
key does not change the GPIB
address.
An advantage of using the *RST command is that it sets
the Agilent 86121A into the single measurement acqui-
sition mode. Because the READ and MEASure data
queries expect this mode, their proper operation is
ensured.
Содержание 86121A
Страница 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Страница 7: ......
Страница 10: ...Chapter 1 Getting Started ...
Страница 21: ......
Страница 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Страница 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Страница 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Страница 32: ...The Softkeys 2 11 The Setup menu ...
Страница 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Страница 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Страница 37: ......
Страница 191: ......
Страница 205: ...7 14 Regulatory Information Declaration of Conformity ...
Страница 239: ......