5
-
50
SENSe Subsystem
Instrument specifications apply when the device is set
to NARRow. Specifications do not apply in BROad
mode.
Query Response
The query form returns the previously selected device.
NARR
ow
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
Sets the elevation value used by the instrument to com-
pensate for air dispersion.
<integer>
is the altitude in meters.
Attribute Summary
Non-sequential command
Preset
State: unaffected by
*RST sets this value to the minimum
SCPI Compliance: instrument specific
Description
Changing the elevation value causes the current data to
be reprocessed.
Query Response
The query form returns the current elevation setting as
shown in the following example:
+1500
:SENSe:CORRection:OFFSet:MAGNitude{?| {<real> | MINimum | MAXimum}}
Enters an offset for amplitude values.
<real>
is the logarithmic units in dB.
Constant
Description
MINimum
0 m
MAXimum
5000 m
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands” on
page 4-11
for more information.
Содержание 86121A
Страница 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Страница 7: ......
Страница 10: ...Chapter 1 Getting Started ...
Страница 21: ......
Страница 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Страница 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Страница 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Страница 32: ...The Softkeys 2 11 The Setup menu ...
Страница 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Страница 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Страница 37: ......
Страница 191: ......
Страница 205: ...7 14 Regulatory Information Declaration of Conformity ...
Страница 239: ......