AMPLitude:LEVel:CIDigits:LEADing
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example
10 OUTPUT 707;”:MEASURE:AMPLITUDE:LEVEL:CIDIGITS:LAGGING 5”
Query
MEASure:AMPLitude:LEVel:CIDigits:LAGGing?
Returned Format
[:MEASure:AMPLitude:LEVel:CIDigits:LAGGing] <num_digits><NL>
AMPLitude:LEVel:CIDigits:LEADing
Command
MEASure:AMPLitude:LEVel:CIDigits:LEADing <num_digits>
Specifies the minimum number of leading consecutive identical digits to be uses in defining
the one and zero levels.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example
10 OUTPUT 707;”:MEASURE:AMPLITUDE:LEVEL:CIDIGITS:LEADING 5”
Query
MEASure:AMPLitude:LEVel:CIDigits:LEADing?
Returned Format
[:MEASure:AMPLitude:LEVel:CIDigits:LEADing] <num_digits><NL>
AMPLitude:LEVel:DEFine
Command
MEASure:AMPLitude:LEVel:DEFine {AVERage | CIDigits}
Specifies whether the one and zero levels are defined as the average one and zero levels, or
whether they are specified in terms of a minimum number of consecutive identical digits.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example
10 OUTPUT 707;”:MEASURE:AMPLITUDE:LEVEL:DEFINE AVERAGE”
Query
MEASure:AMPLitude:LEVel:DEFine?
Returned Format
[:MEASure:AMPLitude:LEVel:DEFine] {AVERage | CIDigits}<NL>
AMPLitude:LOCation
Command
MEASure:AMPLitude:LOCation <location_value>
Specifies the location within the unit interval at which the noise and interference will be mea-
sured. The location is specified as a percentage. 0% corresponds to the left crossing point,
100% corresponds to the right crossing point, and 50% (the default) corresponds to the cen-
ter of the eye. The specified location must be within the range of 5.0% and 95.0%.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example
10 OUTPUT 707;”:MEASURE:AMPLITUDE:LOCATION 25”
Query
MEASure:AMPLitude:LOCation?
Returned Format
[:MEASure:AMPLitude:LOCation] <location_value><NL>
AMPLitude:OLEVel?
Query
MEASure:AMPLitude:OLEVel?
Содержание 86100A
Страница 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Страница 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Страница 94: ...2 40 Sample Programs BASIC Programming Examples...
Страница 124: ...4 16 Root Level Commands VIEW...
Страница 125: ...5 DATE 5 2 DSP 5 2 ERRor 5 3 HEADer 5 4 LONGform 5 5 MODE 5 6 SETup 5 7 TIME 5 7 System Commands...
Страница 132: ...5 8 System Commands TIME...
Страница 140: ...6 8 Acquire Commands SWAVeform RESet...
Страница 152: ...7 12 Calibration Commands STATus...
Страница 192: ...11 12 Display Commands SSAVer...
Страница 202: ...12 10 Function Commands VERTical RANGe...
Страница 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Страница 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Страница 332: ......
Страница 353: ...26 DISPlay 26 2 LOAD 26 2 SAVE 26 3 XOFFset 26 3 XRANge 26 3 YOFFset 26 3 YRANge 26 4 Waveform Memory Commands...