Performance Verification
Purpose
By performing the verification procedure upon delivery of the
86038B, you may have confidence that:
The PDLA is correctly connected
The PDLA was received in working order without damage due
to shipment
Additionally, you may perform this procedure periodically to
verify that the system is still in good operating condition. By
recording the results of this procedure, and tracking these
results over time, you will be able to detect any changes that
may affect the measurement results of the PDLA.
Things to Consider
• Dispersion measurements are very sensitive to temperature
changes. Insure that the PDLA has warmed up for at least 90
minutes. The verification DUT must also be allowed to reach
ambient temperature for at least 60 minutes prior to
measurement.
• Dirty connections will cause poor measurement quality. Be
sure to clean all fiber ends, connectors, and test ports before
making connections. Refer to
.
• This procedure will take approximately 3 hours to complete
(including the required 90 minute warm-up time). Allow
enough time to fully complete the procedure.
• If you need assistance on navigating the front panel
functions, refer to
“Navigating a Setup Panel” on page 46
.
Procedure
1 On the front panel of the PDLA, press
Preset
.
2 Verify optical path. This will assure that the optical signal
can be generated and detected and that good quality optical
connections have been made.
a Connect the 2-meter reference patch-cord (not the MXF)
between the Source Output and Optical Receiver
Содержание 86038B
Страница 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Страница 4: ...4 ...
Страница 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Страница 58: ...58 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 11 86038B Block Diagram ...
Страница 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Страница 230: ...230 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition VB 6 0 Syntax pdlaClient Actions PrintScreen ...
Страница 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Страница 349: ......