85103F Performance Upgrade Installation Manual
25
The 85103F Upgrade Package (8510A/B to 8510C)
Checking Calibration Kit Definitions
Checking Calibration Kit Definitions
After an upgrade, an 8510 contains calibration kit definitions that may be significantly
different from those that were in the pre-upgraded analyzer. Ensure that the calibration
kit definitions contained in the 8510 match the calibration kit used by reloading the
definitions from disk.
Viewing Calibration Kit Labels
1. On the 85102, press
CAL
.
2. Check the calibration kits listed.
3. Load the calibration kit definition for the calibration kit used from disk.
Loading a Calibration Kit Definition
1. Insert the calibration kit data disk into the 8510 disk drive.
The master calibration constants disk (8510 Calibration Data disk, p/n 08510-10034
included in the upgrade package) contains calibration constants for many standard
calibration kits. The calibration kits are listed by model number on the disk label. This
disk can be used to load calibration kit definitions in case the calibration kit does not
have a disk.
2. Press
AUXILIARY MENUS TAPE (or DISC)
and select
[LOAD]
.
3. Select
[CAL KIT 1-2]
.
4. Select the data file to load, normally
FILE 1
.
5. Select the calibration kit you want to replace with the definition from disk.
a. Using the RPG knob or up/down arrow keys, select the appropriate kit.
b. Select
[LOAD FILE]
.
c. Press
[CAL]
to verify that the calibration kit loaded properly.
NOTE
For best measurement results, the 8510 must have the proper
calibration kit definition. If you have any doubt, reload the definition
from disk.