3-14
85056A
Use, Maintenance, and Care of the Devices
Gaging Connectors
6. Note the gage reading. If it is no longer within the allowable range, perform step 4
again.
7. Move the center conductor pullback handle to the unlocked position and then back to
the locked position. The gage reading should return to the value previously set. If not,
repeat steps 4 through 7.
8. Return to
“Gaging the 2.4 mm Sliding Loads” on page 3-11
.
Figure 3-4
Adjusting the Sliding Load Pin Depth