3-10
85054D
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Female Type-N Connectors
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy.
1. Select the proper gage for your connector. (Refer to
for the gage set part
number).
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to
and
earlier in this chapter.
3. Zero the connector gage (refer to
a. While holding the gage by the barrel, and without turning the gage or the gage
master, screw the gage connecting nut onto the female gage master, just until you
meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to
for more information.
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero.
d. Tighten the dial lock screw and remove the gage master.
e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to
a. While holding the gage by the barrel, and without turning the gage or the device,
screw the gage connecting nut onto the device being measured, just until you meet
resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting
nut to 135 N-cm (12 in-lb). Refer to
for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor
is
protruding
by an amount indicated by the
black
numbers. If the needle has moved
counterclockwise, the center conductor is
recessed
by an amount indicated by the
red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take
an average of the readings. After each measurement, rotate the gage a quarter-turn
to reduce measurement variations that result from the gage or the connector face not
being exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in
Содержание 85054D
Страница 5: ...85054D 1 1 1 General Information...
Страница 14: ...1 10 85054D General Information When to Calibrate...
Страница 15: ...2 1 2 Specifications...
Страница 20: ...2 6 85054D Specifications Electrical Specifications...
Страница 21: ...3 1 3 Use Maintenance and Care of the Devices...
Страница 29: ...85054D 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Male Type N Connectors...
Страница 35: ...4 1 4 Performance Verification...
Страница 38: ...4 4 85054D Performance Verification Recertification...
Страница 39: ...5 1 5 Troubleshooting...
Страница 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Страница 45: ...6 1 6 Replaceable Parts...
Страница 48: ...6 4 85054D Replaceable Parts Introduction...
Страница 49: ...A 1 A Standard Definitions...