![Agilent Technologies 85054D Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/agilent-technologies/85054d/85054d_users-and-service-manual_2868611012.webp)
1-8
85054D
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of cable
connectors and adapters is referred to in terms of the center conductor. For example, a
connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
8510-series, 872x, and 875x ONLY:
In contrast, during a measurement calibration, the
network analyzer softkey menus label a 1.85 mm calibration device with reference to the
sex of the analyzer’s test port connector—not the calibration device connector. For
example, the label
SHORT(F)
refers to the short that is to be connected to the female test
port. This will be a male short from the calibration kit.
A connector gage is referred to in terms of the connector that it measures. For instance, a
male connector gage has a female connector on the gage so that it can measure male
devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of the above are described in
Chapter 3 , “Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can
degrade repeatability and accuracy and can damage any connector mated to it. Improper
connections, resulting from pin depth values being out of the
observed
limits (see
), or from bad connections, can also damage these devices.
Table 1-2 Clarifying the Terminology of Connectors: Examples
Terminology
Meaning
Short
–f–
Female short (female center conductor)
Short (f)
Male short (male center conductor) to be connected to female port
Содержание 85054D
Страница 5: ...85054D 1 1 1 General Information...
Страница 14: ...1 10 85054D General Information When to Calibrate...
Страница 15: ...2 1 2 Specifications...
Страница 20: ...2 6 85054D Specifications Electrical Specifications...
Страница 21: ...3 1 3 Use Maintenance and Care of the Devices...
Страница 29: ...85054D 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Male Type N Connectors...
Страница 35: ...4 1 4 Performance Verification...
Страница 38: ...4 4 85054D Performance Verification Recertification...
Страница 39: ...5 1 5 Troubleshooting...
Страница 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Страница 45: ...6 1 6 Replaceable Parts...
Страница 48: ...6 4 85054D Replaceable Parts Introduction...
Страница 49: ...A 1 A Standard Definitions...