background image

4-2

85052D

 

Performance Verification

Introduction

Introduction

The performance of your calibration kit can only be verified by returning the kit to Agilent 
Technologies for recertification. The equipment required to verify the specifications of the 
devices in the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in Your Kit

Agilent verifies the specifications of these devices as follows:

1. The residual microwave error terms of the test system are verified with precision 

airlines and shorts that are directly traced to the National Institute of Standards and 
Technology (NIST). The airline and short characteristics are developed from mechanical 
measurements. The mechanical measurements and material properties are carefully 
modeled to give very accurate electrical representation. The mechanical measurements 
are then traced to NIST through various plug and ring gages and other mechanical 
measurements.

2. Each calibration device is electrically tested on this system. For the initial (before sale) 

testing of the calibration devices, Agilent includes the test measurement uncertainty as 
a guardband to guarantee each device meets the published specification. For 
recertifications (after sale), no guardband is used and the measured data is compared 
directly with the specification to determine the pass or fail status. The measurement 
uncertainty for each device is, however, recorded in the calibration report that 
accompanies recertified kits.

These two steps establish a traceable link to NIST for Agilent to the extent allowed by the 
institute’s calibration facility. The specifications data provided for the devices in the kit is 
traceable to NIST through Agilent Technologies.

Содержание 85052D

Страница 1: ...ration Kit This manual directly applies to 85052D calibration kits with serial number prefix 3106A Agilent Part Number 85052 90079 Printed in USA Print Date November 2004 Supersedes March 2004 Copyrig...

Страница 2: ...CUMENT THAT CONFLICT WITH THESE TERMS THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL DFARS Restricted Rights Notice If software is for use in the performance of a U S Government prime contr...

Страница 3: ...ntation from the Web download the PDF file from the Agilent web site Go to http www agilent com Enter the document s part number located on the title page in the Quick Search box Click GO Click on the...

Страница 4: ...iv...

Страница 5: ...5 Residual Errors after Calibration 2 6 Certification 2 6 3 Use Maintenance and Care of the Devices Electrostatic Discharge 3 2 Visual Inspection 3 3 Look for Obvious Defects and Damage First 3 3 Ins...

Страница 6: ...e Information 5 3 Contacting Agilent 5 4 6 Replaceable Parts Introduction 6 2 A Standard Definitions Standard Class Assignments A 2 Blank Forms A 5 Nominal Standard Definitions A 8 Setting the System...

Страница 7: ...85052D 1 1 1 General Information...

Страница 8: ...to Chapter 6 Replaceable Parts for a complete list of kit contents and their associated part numbers Broadband Loads The broadband loads are metrology grade terminations that have been optimized for...

Страница 9: ...n the kit must be installed in the analyzer if not permanently stored in the internal memory or hard drive The calibration definitions can be recalled from the analyzer if stored in internal memory or...

Страница 10: ...n page 5 3 Recording the Device Serial Numbers In addition to the kit serial number the devices in the kit are individually serialized serial numbers are labeled onto the body of each device Record th...

Страница 11: ...ng the Sex of a Connector In this manual and in the prompts of the PNA calibration wizard the sex of calibration devices and adapters is referred to in terms of the calibration device connector For ex...

Страница 12: ...to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it Improper connections resulting from pin depth...

Страница 13: ...85052D 2 1 2 Specifications...

Страница 14: ...quirements Parameter Limits Temperature Operatinga a The temperature range over which the calibration standards maintain conformance to their specifications 20 C to 26 C Storage 40 C to 75 C Error cor...

Страница 15: ...age 3 6 explains how to use gages to determine if the kit devices have maintained their mechanical integrity Refer to Table 2 2 for typical and observed pin depth limits Pin Depth Pin depth is the dis...

Страница 16: ...in depth slightly recessed Table 2 2 Pin Depth Limits Device Typical Pin Depth Measurement Uncertaintya a Approximately 2 sigma to 2 sigma of gage uncertainty based on studies done at the factory acco...

Страница 17: ...urn loss 34 dB 0 01995 20 to 26 5 Offset opensa a The specifications for the opens and shorts are given as allowed deviation from the nominal model as defined in the standard definitions see Nominal S...

Страница 18: ...ormation on how to use the software Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory Agilent further certifies t...

Страница 19: ...85052D 3 1 3 Use Maintenance and Care of the Devices...

Страница 20: ...aking connections always wear a heel strap when working in an area with a conductive floor If you are uncertain about the conductivity of your floor wear a heel strap always ground yourself before you...

Страница 21: ...re not connected properly Connector wear eventually degrades performance of the device Calibration devices should have a long life if their use is on the order of a few times per week Replace devices...

Страница 22: ...precision measurements with conventional slotted connectors The precision slotless connector was developed to eliminate this problem The PSC has a center conductor with a solid cylindrical shell the...

Страница 23: ...connector that cannot be removed with compressed air or nitrogen Refer to Table 6 2 on page 6 3 for part numbers for isopropyl alcohol and cleaning swabs a Apply a small amount of isopropyl alcohol to...

Страница 24: ...sultant average can be in error by as much as 0 0001 inch due to systematic biasing errors usually resulting from worn gages and gage masters The information in Table 2 2 on page 2 4 assumes new gages...

Страница 25: ...ure readings It will serve as a good troubleshooting tool when you suspect damage may have occurred to the device If either visual inspection or electrical performance suggests that the connector inte...

Страница 26: ...formation c As you watch the gage pointer gently tap the barrel of the gage to settle the reading The gage pointer should line up exactly with the zero mark on the gage If not adjust the zero set knob...

Страница 27: ...85052D 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 3 5 mm Connectors...

Страница 28: ...n the connector nut finger tight CAUTION Do not turn the device body Only turn the connector nut Damage to the center conductor can occur if the device body is twisted Do not twist or screw the connec...

Страница 29: ...nches within 90 degrees of each other before applying force See Figure 3 3 Wrenches opposing each other greater than 90 degrees apart will cause a lifting action which can misalign and stress the conn...

Страница 30: ...vice body is twisted 1 Use an open end wrench to prevent the device body from turning 2 Use another open end wrench to loosen the connecting nut 3 Complete the separation by hand turning only the conn...

Страница 31: ...85052D 4 1 4 Performance Verification...

Страница 32: ...s are carefully modeled to give very accurate electrical representation The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements 2 Each...

Страница 33: ...to Contacting Agilent on page 5 4 for a list of offices How Often to Recertify The suggested initial interval for recertification is 12 months or sooner The actual need for recertification depends on...

Страница 34: ...4 4 85052D Performance Verification Recertification...

Страница 35: ...85052D 5 1 5 Troubleshooting...

Страница 36: ...oting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration or if your network analyzer does not pass performance verification follow the steps in Figure 5 1 Figure 5 1 Trou...

Страница 37: ...ress a technical contact person within your company and the person s complete telephone number the model number and serial number of the kit the part number and serial number of each device the type o...

Страница 38: ...20 87 44 11 fax 0820 87 44 22 Belgium tel 32 0 2 404 9340 alt 32 0 2 404 9000 fax 32 0 2 404 9395 Denmark tel 45 7013 1515 alt 45 7013 7313 fax 45 7013 1555 Finland tel 358 10 855 2100 fax 358 0 10 85...

Страница 39: ...85052D 6 1 6 Replaceable Parts...

Страница 40: ...e Parts for the 85052D Calibration Kit Description Qty per kit Agilent Part Number Calibration Devices 3 5 mm Broadband load m a 1 00902 60003 Broadband load f a 1 00902 60004 Offset short m 1 85052 6...

Страница 41: ...SD Protection Devices Grounding wrist strap 1 9300 1367 5 ft grounding cord for wrist strap 1 9300 0980 2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300 0797 ESD heel strap 1 9300 13...

Страница 42: ...6 4 85052D Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85052D Calibration Kit...

Страница 43: ...85052D 6 5 Replaceable Parts Introduction Figure 6 2 Replaceable Parts for the 85052D Calibration Kit...

Страница 44: ...6 6 85052D Replaceable Parts Introduction...

Страница 45: ...85052D A 1 A Standard Definitions...

Страница 46: ...rd Class Assignments for the 8510 Network Analyzer Disk File Name CK_35mmD3 Calibration Kit Label 3 5 mm D 3 Class A B C D E F G Standard Class Label S11A 2 Open S11B 1 Short S11C 9 Loads S22A 2 Open...

Страница 47: ...l S11A 2 Open S11B 1 Short S11C 3 Loads S22A 2 Open S22B 1 Short S22C 3 Loads Forward transmission 4 Thru Reverse transmission 4 Thru Forward match 4 Thru Reverse match 4 Thru Response 1 2 4 Response...

Страница 48: ...defined as match standards 3 If you are performing an LRM calibration S21T and S12T must be defined as line standards S11A and S22A must be defined as reflection standards S11B S11C S22B and S22C must...

Страница 49: ...Kit Label _________________ Class A B C D E F G Standard Class Label S11A S11B S11C S22A S22B S22C Forward transmission Reverse transmission Forward match Reverse match Forward isolationa a The forwar...

Страница 50: ...C D E F G Standard Class Label S11A S11B S11C S22A S22B S22C Forward transmission Reverse transmission Forward match Reverse match Response Response isolation TRL thru TRL reflect TRL line or match T...

Страница 51: ...st be defined as match standards 3 If you are performing an LRM calibration S21T and S12T must be defined as line standards S11A and S22A must be defined as reflection standards S11B S11C S22B and S22...

Страница 52: ...ecify the mathematical model of each device This information must be loaded into the network analyzer to perform valid calibrations Refer to your network analyzer user s guide for instructions on load...

Страница 53: ...ax Open 3 4 5 Openf 6 9558 1 0259 0 01435 0 0028 0 50 0 0 999 Coax 3 5 2 92 6 Openf 5 9588 11 195 0 5076 0 00243 0 50 0 0 999 Coax 3 5 SMA 7 Openf 13 4203 1 9452 0 5459 0 01594 0 50 0 0 999 Coax 2 92...

Страница 54: ...or Waveguide Standard Label Number Type Delay in ps Z 0 in Loss in G s Min Max 1 Short 0 0 0 0 31 798 50 2 191 0 999 Coax Short 2 Open 49 43 310 13 23 17 0 16 29 243 50 2 2 0 999 Coax Open 3 Load Fxd...

Страница 55: ...de Standard Label Number Type L0 10 12 H L1 10 24 H Hz L2 10 33 H Hz 2 L3 10 45 H Hz 3 Delay in ps Z 0 Loss in G s Min Max 1 Short 2 0765 108 54 2 1705 0 01 31 785 50 2 366 0 999 Coax Short 2 Open 49...

Страница 56: ...Ensure system Z0 of network analyzer is set to this value Calibration Kit Label ____________________________ Standardb b Open short load delay thru or arbitrary impedance C0 10 15 F C1 10 27 F Hz C2...

Страница 57: ...C1 10 27 F Hz C2 10 36 F Hz 2 C3 10 45 F Hz 3 Fixed c or Sliding Offset Frequency in GHzd Coax or Waveguide Standard Label Number Type Delay in ps Z 0 in Loss in G s Min Max 1 2 3 4 5 6 7 8 a Ensure...

Страница 58: ...Hz C2 10 36 F Hz 2 C3 10 45 F Hz 3 Fixed or sliding Offset Frequency in GHzc Coax or Waveguide Standard Label Number Type L0 10 12 H L1 10 24 H Hz L2 10 33 H Hz 2 L3 10 45 H Hz 3 Delay in ps Z 0 Loss...

Страница 59: ...r cleaning 3 4 damage 3 3 defects 3 3 female 3 4 gage accuracy 3 6 handling 3 6 3 8 use of 3 6 zeroing 3 6 3 8 gaging 3 6 3 8 to determine pin depth 3 6 when to do 3 7 mating plane surfaces cleaning 3...

Страница 60: ...cteristics 2 3 affect on electrical performance 2 3 verifying 3 6 MIL STD 45662 calibration 4 3 N National Institute of Standards and Technology NIST 2 6 4 2 nitrogen for cleaning 3 4 nominal standard...

Страница 61: ...and Technology NIST 2 6 4 2 static discharge 3 2 electricity 3 2 storage 3 12 storage case part number 6 2 strap heel and wrist part number 6 3 supplies cleaning 1 3 part number 6 3 swabs cleaning 3...

Страница 62: ...Index iv 85052D Index...

Отзывы: