![Agilent Technologies 85024A Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925037.webp)
Chapter 5
5-7
Performance Tests
Gain Compression
16. Connect the probe, the probe adapter, and a barrel between the attenuator and the power sensor. The
signal from the analyzer should now go through the attenuator and then the probe before reaching the
power sensor. Record the power reading on the
(item F).
17. Remove the attenuator pad, connect the probe adapter to the output port of the analyzer. Connect the
attenuator pad between the probe and the power sensor. The signal from the analyzer should now go
through the probe and then the pad before reaching the power sensor. Record the power reading on
the
(item G).
18. Perform the calculation indicated on the test record to determine the result of the test.
Calculating Equivalent Power
The probe is usually used as a high impedance device. The compression specification is given as a
voltage (instead of power) because of the wide range of devices with which the probe may be used. For
testing the probe, the probe adapter is used to make the probe input 50
. Using this known impedance,
an equivalent power can be calculated.
To find the power level equivalent to the 0.3 peak voltage, perform the following mathematical steps:
Steps
Formula
Example
1. Convert the peak voltage into RMS.
2. Convert the RMS voltage to a raw power.
3. Convert the raw power to a power
which is relative to 1 mW.
divide by 0.001w
4. Convert to dBm.
* Therefore, as shown in the example, the equivalent power to 0.3 V peak in a 50
system is
0.458
dBm.
V
pk
1.4142135
-------------------------
0.3
V
1.4142135
-------------------------
0.212132
V
R MS
=
V
2
R
------
0.212132
2
50
-----------------------------
0.0009
w
=
0.009
w
0.001
w
-----------------
0.9
=
10 LOGx
10 LOG 0.9
0.458 dBm
*
–
=
Содержание 85024A
Страница 4: ...iv ...
Страница 7: ...1 1 1 General Information ...
Страница 13: ...2 1 2 Accessories ...
Страница 19: ...3 1 3 Installation ...
Страница 24: ...3 6 Chapter3 Installation Returning the Product for Service ...
Страница 25: ...4 1 4 Operation ...
Страница 30: ...4 6 Chapter4 Operation Operator s Check ...
Страница 31: ...5 1 5 Performance Tests ...
Страница 40: ...5 10 Chapter5 Performance Tests Average Noise Level ...
Страница 41: ...6 1 6 Replaceable Parts ...
Страница 46: ...6 6 Chapter6 Replaceable Parts Parts Lists ...
Страница 47: ...7 1 7 Service ...
Страница 60: ...7 14 Chapter7 Service Replacement Procedure Figure 7 11 Regulator Parts and Wiring ...
Страница 64: ...7 18 Chapter7 Service Connector Inspection and Cleaning ...