1-2
Installation
Installation
Installation
Agilent 83491/2/3/4A modules require that firmware revision A.06.25 or later
be installed in the Agilent 83480A. If you wish to install the module in an
Agilent 54750A digitizing oscilloscope, you must first install the
Agilent 83480K communications firmware upgrade kit.
N O T E
The front panel of Agilent 83494A clock recovery modules have an auxiliary clock
output that provides the recovered clock signal,
refer to “Auxiliary outputs” on page 2-6
.
To use this output with the 9.953 and 10.664 Gb/s (Option 106) data rates, the digital
communications analyzer mainframe with the 12 GHz (Divided Trigger) option is
required.
To check the Agilent 83480A’s firmware revision code
1
Press the
Utility
key and then the
System config
softkey.
2
The firmware revision number is listed under the Frame section of the display.
C A U T I O N
Fiber-optic connectors are easily damaged when connected to dirty or
damaged cables and accessories. The Agilent 83492A, 83493A, and 83494A
front-panel input connectors are no exception. When you use improper
cleaning and handling techniques, you risk expensive instrument repairs,
damaged cables, and compromised measurements. Before you connect any
fiber-optic cable to an Agilent 83492A, 83493A, or 83494A module,
“Cleaning Connections for Accurate Measurements” on page 4-10
.
C A U T I O N
The circuits on electrical inputs and outputs can be damaged by electrostatic
discharge (ESD). Therefore, avoid applying static discharges to any front or
rear-panel electrical connector. Before connecting any coaxial cable to a front-
panel connector, momentarily short the center and outer conductors of the
cable together. Avoid touching the front-panel connectors without first
touching the frame of the instrument. Be sure that the instrument is properly
earth-grounded to prevent buildup of static charge.
Discharge Information” on page 4-8
book.book Page 2 Monday, January 7, 2002 2:12 PM
Содержание 83491
Страница 4: ...iv General Safety Considerations book book Page iv Monday January 7 2002 2 12 PM ...
Страница 6: ...book book Page 2 Monday January 7 2002 2 12 PM ...
Страница 14: ...book book Page 8 Monday January 7 2002 2 12 PM ...
Страница 30: ...2 16 Operation Using Probes with an Agilent 83491A book book Page 16 Monday January 7 2002 2 12 PM ...
Страница 66: ...book book Page 24 Monday January 7 2002 2 12 PM ...