Reference
Electrostatic Discharge Information
Both types, when used together, provide a significant level of ESD protection.
Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 M
Ω
of
isolation from ground. Refer to
for information on ordering static-
safe accessories.
W A R N I N G
These techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than 500
volts.
Reducing ESD Damage
The following suggestions may help reduce ESD damage that occurs during
testing and servicing operations.
• Personnel should be grounded with a resistor-isolated wrist strap before re-
moving any assembly from the unit.
• Be sure all instruments are properly earth-grounded to prevent a buildup of
static charge.
Table 4-1. Static-Safe Accessories
Agilent Part
Number
Description
9300-0797
Set includes: 3M static control mat 0.6 m
×
1.2 m (2 ft
×
4 ft) and 4.6 cm (15
ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They
must be ordered separately.)
9300-0980
Wrist-strap cord 1.5 m (5 ft)
9300-1383
Wrist-strap, color black, stainless steel, without cord, has four adjustable
links and a 7 mm post-type connection.
9300-1169
ESD heel-strap (reusable 6 to 12 months).
book.book Page 9 Monday, January 7, 2002 2:12 PM
Содержание 83491
Страница 4: ...iv General Safety Considerations book book Page iv Monday January 7 2002 2 12 PM ...
Страница 6: ...book book Page 2 Monday January 7 2002 2 12 PM ...
Страница 14: ...book book Page 8 Monday January 7 2002 2 12 PM ...
Страница 30: ...2 16 Operation Using Probes with an Agilent 83491A book book Page 16 Monday January 7 2002 2 12 PM ...
Страница 66: ...book book Page 24 Monday January 7 2002 2 12 PM ...