4-3
Specifications and Regulatory Information
Specifications
Specifications
Table 4-1. Agilent 83487A Electrical Channel Vertical Specifications
Bandwidth (–3 dB)
dc Accuracy—single voltage marker
a
dc to 12.4 GHz or 20 GHz, user selectable
12.4 GHz
±
0.4% of full scale
±
2 mV
±
1.5% (reading – channel offset)
±
(2%/
°
C) (
∆
T
cal
b
) (reading) – 0.4%/hr (
∆
Time
cal
c
) (reading)
20 GHz
±
0.4% of full scale
±
2 mV
±
3% of reading– channel offset
±
(2%/
°
C) (
∆
T
cal
b
) (reading) – 0.4%/hr (
∆
Time
cal
c
) (reading)
dc Difference—two marker accuracy on
same channel
a
12.4 GHz
±
0.8% of full scale
±
1.5% of delta marker reading
±
(2%/
°
C) (
∆
T
cal
b
) (reading) – 0.4%/hr (
∆
Time
cal
c
) (reading)
20 GHz
±
0.8% of full scale
±
3% of delta marker reading
±
(2%/
°
C) (
∆
T
cal
b
) (reading) – 0.4%/hr (
∆
Time
cal
c
) (reading)
Transition Time (10% to 90%)
calculated from T=0.35/BW,
characteristic
12.4 GHz
≤
28.2 ps
20 GHz
≤
17.5 ps
Maximum RMS Noise
12.4 GHz
≤
0.5 mV (0.25 mV typical)
Содержание 83487A
Страница 1: ...Agilent 83487A Optical Electrical Plug In Module User s Guide ...
Страница 8: ...viii General Safety Considerations ...
Страница 10: ......
Страница 25: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Страница 27: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Страница 28: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Страница 38: ......
Страница 67: ...4 9 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 68: ......
Страница 96: ......