5-33
Servicing
Replacing Instrument Assemblies
To replace line filter assembly FL1
Although replacing FL1 is straightforward, be sure to observe the following
points:
• Tighten the screws that attach FL1 to 6 in-lbs. Do not use the recommended
torque for M3 screws. Over tightening these screws will damage the line filter’s
flange.
• Observe the following warning about correct wire attachment. The wire color
codes in the figure refer to the following cable colors:
54 green/yellow
98 white/gray
918 white/brown/gray
W A R N I N G
Be sure to solder the wires to FL1 in the correct positions as
shown in the following figure. Failure to attach these wires
correctly could result in damage to the instrument and injury
to the user.
Содержание 83437A
Страница 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Страница 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Страница 8: ......
Страница 10: ......
Страница 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Страница 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Страница 47: ...3 7 Specifications and Regulatory Information Regulatory Information Declaration of Conformity I ll ...
Страница 48: ......
Страница 54: ......
Страница 61: ...5 7 Servicing General Information ...
Страница 63: ...5 9 Servicing General Information ...
Страница 79: ...5 25 Servicing Adjustment Procedure ...
Страница 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Страница 92: ...5 38 Servicing Replaceable Parts ...
Страница 94: ...5 40 Servicing Replaceable Parts ...
Страница 96: ...5 42 Servicing Replaceable Parts ...
Страница 98: ...5 44 Servicing Replaceable Parts ...
Страница 100: ...5 46 Servicing Replaceable Parts ...
Страница 106: ......