5-20
Servicing
Performance Tests
Performance Tests
The procedures in this section test the Agilent 83438A’s performance using
the specifications listed in Chapter 3, “Specifications and Regulatory Informa-
tion” as the performance standard. All of the tests are done manually without
the aid of a computer. None of these tests require access to the interior of the
instrument. Allow the Agilent 83438A to warm up for 15 minutes before doing
any of the performance tests.
Calibration Cycle
This instrument requires periodic verification of performance. The instrument
should have a complete verification of specifications at least once every two
years.
C A U T I O N
Option 022 instruments have an angled-fiber output. Be sure to use an angled-
fiber patchcord during testing. If you do not have an angled-fiber patchcord
available, you can purchase the required accessories from Agilent
Technologies. These include the Agilent 81000SI DIN connector interface and
the Agilent 81113PC DIN 4108 (angled) to Super-PC patchcord.
Содержание 83437A
Страница 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Страница 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Страница 8: ......
Страница 10: ......
Страница 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Страница 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Страница 47: ...3 7 Specifications and Regulatory Information Regulatory Information Declaration of Conformity I ll ...
Страница 48: ......
Страница 54: ......
Страница 61: ...5 7 Servicing General Information ...
Страница 63: ...5 9 Servicing General Information ...
Страница 79: ...5 25 Servicing Adjustment Procedure ...
Страница 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Страница 92: ...5 38 Servicing Replaceable Parts ...
Страница 94: ...5 40 Servicing Replaceable Parts ...
Страница 96: ...5 42 Servicing Replaceable Parts ...
Страница 98: ...5 44 Servicing Replaceable Parts ...
Страница 100: ...5 46 Servicing Replaceable Parts ...
Страница 106: ......