5-36
Servicing
Adjustment Procedures
Adjustment Procedures
Perform the adjustments in this section if the instrument fails any of its per-
formance tests. Periodic adjustment is
not
required to maintain safety.
The following adjustment procedures are provided in this section:
To remove the instrument cover
Adjustment 2. Laser Bias and Control
5-41
W A R N I N G
Improper internal adjustments may change the laser classification of
this instrument. Always perform the adjustments exactly as described
in this section.
C A U T I O N
Option 022 instruments have an angled-fiber output. Be sure to use an angled-
fiber patchcord during testing. If you do not have an angled-fiber patchcord
available, you can purchase the required accessories from Agilent
Technologies. These include the Agilent 81000SI DIN connector interface and
the Agilent 81113PC DIN 4108 (angled) to Super-PC patchcord.
W A R N I N G
Do not remove or relocate any shunt jumpers located on any printed
circuit assembly.
Содержание 83430A
Страница 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Страница 9: ......
Страница 11: ......
Страница 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Страница 25: ......
Страница 31: ...2 6 Making Measurements Using the Agilent 83430A...
Страница 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Страница 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Страница 51: ......
Страница 63: ...5 6 Servicing General Information...
Страница 69: ...5 12 Servicing Troubleshooting...
Страница 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Страница 103: ......