5-22
Servicing
Performance Tests
Test 4. Electrical Return Loss (DIGITAL IN)
1
Press the
PRESET
key on the Agilent 8752C network analyzer.
2
With averaging on, calibrate the Agilent 8752C for a 1-port reflection
measurement at the end of the RF output cable. Use the calibration termination
kit. A through calibration is not required.
3
Connect the equipment as shown in the following figure.
Figure 5-6. Equipment setup
4
Press the Agilent 83430A’s front-panel
SELECT
key so that the
DIGITAL IN AC
COUPLED
light is on.
5
Restart averaging on the Agilent 8752C.
6
Confirm that the return loss meets the following requirements:
≤
1.0 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . >12.07 dB
1 GHz to 2 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . >9.07 dB
2 GHz to 2.5 GHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . >6.07 dB
7
Disconnect the RF cable from the
DIGITAL IN AC COUPLED
connector, and
reconnect the cable to the
DIGITAL IN DC COUPLED
connector.
8
Press the Agilent 83430A’s front-panel
SELECT
key so that the
DIGITAL IN DC
COUPLED
light is on.
9
Restart averaging on the Agilent 8752C.
Содержание 83430A
Страница 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Страница 9: ......
Страница 11: ......
Страница 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Страница 25: ......
Страница 31: ...2 6 Making Measurements Using the Agilent 83430A...
Страница 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Страница 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Страница 51: ......
Страница 63: ...5 6 Servicing General Information...
Страница 69: ...5 12 Servicing Troubleshooting...
Страница 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Страница 103: ......