2-16
Making Measurements
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections
on the cable end face can be detected as well as cracks or chips in the fiber
itself. Use a microscope (100X to 200X magnification) to inspect the entire
end face for contamination, raised metal, or dents in the metal as well as any
other imperfections. Inspect the fiber for cracks and chips. Visible imperfec-
tions not touching the fiber core may not affect performance (unless the
imperfections keep the fibers from contacting).
W A R N I N G
Always remove both ends of fiber-optic cables from any instrument,
system, or device before visually inspecting the fiber ends. Disable all
optical sources before disconnecting fiber-optic cables. Failure to do
so may result in permanent injury to your eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber-
optic cables and Agilent Technologies universal adapters. The initial cleaning,
using the alcohol as a solvent, gently removes any grit and oil. If a caked-on
layer of material is still present, (this can happen if the beryllium-copper sides
of the ferrule retainer get scraped and deposited on the end of the fiber during
insertion of the cable), a second cleaning should be performed. It is not
uncommon for a cable or connector to require more than one cleaning.
C A U T I O N
Agilent Technologies strongly recommends that index matching compounds
not
be applied to their instruments and accessories. Some compounds, such as
gels, may be difficult to remove and can contain damaging particulates. If you
think the use of such compounds is necessary, refer to the compound
manufacturer for information on application and cleaning procedures.
Table 2-1. Cleaning Accessories
Item
Agilent Part Number
Pure isoporpyl alcohol
—
Cotton swabs
8520-0023
Small foam swabs
9300-1223
Compressed dust remover (non-residue)
8500-5262
Содержание 83430A
Страница 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Страница 9: ......
Страница 11: ......
Страница 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Страница 25: ......
Страница 31: ...2 6 Making Measurements Using the Agilent 83430A...
Страница 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Страница 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Страница 51: ......
Страница 63: ...5 6 Servicing General Information...
Страница 69: ...5 12 Servicing Troubleshooting...
Страница 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Страница 103: ......