4-8
Performance Tests
Tests for Receivers
3
Use the Agilent 8702D’s “Guided Setup” feature to perform an electrical-to-
electrical response test. Use the following settings:
Start frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Stop frequency (Agilent 83410C/2B): . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 GHz
Stop frequency (Agilent 83411C/D): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 GHz
Stimulus power: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dBm
IF bandwidth: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz
4
The guided setup routine will step you through an S11, 1-port, electrical
calibration.
5
Place the Agilent 8702D in single-sweep mode, and use the marker functions
to locate the peak of the response.
6
Compare the measurement with the specification listed in
“Specifications and Regulatory Information”
.
3. Dynamic Accuracy
Dynamic accuracy is not a specified performance parameter. However, this
procedure is provided for comparison with the characteristic value.
1
Connect the system as shown in
.
2
Turn on the equipment, and allow them to warm up for 1 hour.
3
Set the Agilent 8702D to the following settings:
Start frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Stop frequency:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 GHz
Stimulus power: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
IF bandwidth: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Hz
4
When adjusting the power level in the previous step, set the optical attenuator
so that the maximum E/E trace shown on the Agilent 8702D is –5 dBm
(±0.2 dBm). This is the 0 dBm optical attenuation level.
5
Set the optical attenuator to –10 dB (optical) below the 0 dB level. On the
Agilent 8702D, read the E/E trace as 20 dB below the trace level set in the
previous step.
Содержание 83400 Series
Страница 2: ...Agilent 83400 Series Lightwave Source and Receiver Modules User s Guide ...
Страница 6: ...v Laser Classification IEC and FDA labels and their placement on an Agilent 83402C ...
Страница 9: ......
Страница 11: ......
Страница 17: ...1 6 Getting Started Connecting a Source Figure 1 2 Location of laser safety labels ...
Страница 47: ...2 24 General Information Theory of Operation Figure 2 13 Agilent 83410C and 83412B block diagram ...
Страница 66: ...3 15 Specifications and Regulatory Information Regulatory Information Declaration of conformity sources ...
Страница 67: ...3 16 Specifications and Regulatory Information Regulatory Information Declaration of conformity receivers ...
Страница 81: ...Index 4 Index width spectral procedure 4 4 Y YAG system dual hetrodyne 4 7 ...