Chapter 4, GPIB Commands

Introduction

Common commands such as 

*RST

 may not be used in a subsystem. Common 

commands function regardless of the current specified path. 

Figure 27

 The Correct Use of Colons and Semicolons

Command Abbreviations 

Abbreviations can be used for some commands. This manual does not use 
abbreviations for commands when describing them. However, the letters that are 
used in the abbreviation of the command appear in uppercase, and the letters that 
must be omitted to form the abbreviation appear in lowercase. 

For example, this manual indicates that the abbreviation of the command 

:SYSTEM

 is 

:SYST

 by showing it as 

:SYSTem

. (Abbreviations such as 

:SYSTE

 

are not valid.) 

Returns the current path to root.

Causes the current path to branch one level down.

Does not change the specification of the current path.

Содержание 83236B

Страница 1: ...Manual Serial Numbers This manual applies directly to instruments with serial number prefix 3711J For additional important information about serial numbers see Serial Numbers on page 144 Agilent Part...

Страница 2: ...ins proprietary information that is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translation to another language without the prior written c...

Страница 3: ...changes when a new edition is printed Minor corrections and updates that are incorporated at reprint do not cause the date to change The manual part number changes when extensive technical changes are...

Страница 4: ...s returned to Agilent Technologies from another country Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its prog...

Страница 5: ...strument To avoid electric shock hazard the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade DO NOT Operate In An Explosive Atmosp...

Страница 6: ...finitions of safety symbols used on equipment or in manuals are listed below Instruction manual symbol the product is marked with this symbol when it is necessary for the user to refer to the instruct...

Страница 7: ...vii...

Страница 8: ...tellerbescheinigung GERXUSCHEMISSION LpA 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T 19 Manufacturer s Declaration ACOUSTIC NOISE EMISSION LpA 70 dB operator position normal operation per...

Страница 9: ...escribes how to control the PCS Interface using GPIB commands Appendix 4 GPIB Commands This chapter describes the remote operation of the PCS Interface using GPIB Appendix 5 Specifications This chapte...

Страница 10: ...x...

Страница 11: ...led via Serial Port for CDMA Subscriber Unit Tests 34 When Used with Agilent 8924C and Controlled via GPIB for CDMA Subscriber Unit Tests 37 When Used with Agilent 8920B and Controlled via Serial Port...

Страница 12: ...trol Introduction 66 Overview of Programming Issues 67 PCS Interface Programming Guide 75 Test Procedure Flow 76 Initialize the System 77 Set Up and Measure 79 4 GPIB Commands Introduction 92 The Synt...

Страница 13: ...5 Analyzer Input Path 128 Reference Specifications 131 Remote Control 131 Connectors 132 General Characteristics 135 A Frequency Conversion Tables Frequency Conversion for Signal Generation 138 B Manu...

Страница 14: ...Contents 14 Error Message 154 Glossary 161...

Страница 15: ...15 1 G eneral Information What s included in this chapter Introduction Furnished Accessories Options Front Panel Rear Panel General Information...

Страница 16: ...1A and 83204A composes the Agilent 8921A option 500 2 Can be used with Agilent 8920B only The combination of the Agilent 8920B and 83206A composes the Agilent 8920B option 800 3 The combination of the...

Страница 17: ...User s Guide or Agilent 8920B User s Guide for the operation of the Agilent 83236B controlled via the serial port Conventions Used The Agilent 83236B PCS Interface can be used with a variety of Test S...

Страница 18: ...the contents of the shipment have been checked for completeness and the PCS Interface has been checked mechanically and electrically The contents of the shipment should be as listed in table 2 on page...

Страница 19: ...0 5 m 1 Agilent 10833D Serial Cable 1 Agilent p n 83236 61609 N BNC Cable 2 Agilent p n 83236 61603 SMA SMC Cable 1 Agilent p n 83215 61643 SMC SMC Cable 1 Agilent p n 83215 61644 BNC BNC Cable 1 Agi...

Страница 20: ...instrument is equipped with a three wire power cable When connected to an appropriate ac power outlet this cable grounds the instrument frame The type of power cable shipped with each instrument depen...

Страница 21: ...on Power Cable WARNING For protection from electrical shock the power cable ground must not be defeated The power plug must be plugged into an outlet that provides a protective earth ground connection...

Страница 22: ...m BNC f adapter are also provided for the Connectivity Test Program Option 002 Utility Software on EPSON Memory Card Provides the Utility Software memory cards which include the Manual Control Connect...

Страница 23: ...how to rack mount the PCS Interface refer to the Agilent 83236A B PCS Interface Option AX4 Rack Flange Kit Rack Mount Installation Instructions included with the kit Table 4 Included Accessories Opti...

Страница 24: ...witch The On position is and the Off position is In the On position power is supplied and the indicator above the switch is lit 2 FROM DUPLEX OUT Connects to the DUPLEX OUT port of a Test Set Connecto...

Страница 25: ...ternal attenuator if you are uncertain Subscriber Unit Test Single carrier TDMA and FM 10 Watts CDMA 5 Watts Base Station Test Single carrier TDMA and FM 10 Watts CDMA 1 Watt CAUTION Peak instantaneou...

Страница 26: ...PORT is used to control the PCS Interface 2 REF IN Reference frequency input port The REF IN port connects to the 10 MHz reference source Connector type BNC F 3 REF OUT Reference frequency output port...

Страница 27: ...ar panel This interface conforms to IEEE 488 2 and has the following functions Table 7 GPIB Functions Codes Functions SH1 Full source handshake capability AH1 Full accepter handshake capability T8 Bas...

Страница 28: ...ve or Agilent 8920B whose firmware revision is B 05 00 and above NOTE To enable this serial port set the GPIB Ser switch of the GPIB Address Selector to the Ser side right side as viewed from the rear...

Страница 29: ...PCS Interface The type and rating of this fuse are as follows UL CSA type time delay 3 A 250 Vac WARNING The main fuse is not replaceable by the operator Refer servicing to qualified personnel Ventila...

Страница 30: ...30 Chapter 1 General Information Other Information...

Страница 31: ...ate programs using the PCS Interface it is a good idea to follow the steps in this chapter This will ensure that the proper connections are made between the PCS Interface and other hardware and will v...

Страница 32: ...CS Interface using Refer to CDMA Subscriber Units Agilent 8924C Firmware A 05 00 and above Serial Port page 34 Agilent 8924C Firmware Below A 05 00 GPIB page 37 TDMA Subscriber Units Agilent 8920B Fir...

Страница 33: ...he PCS Interface independently GPIB Control All of the above Test Sets can control the PCS Interface via the GPIB You can use the Manual Control Software or you can make programs to control the PCS In...

Страница 34: ...cessories Follow these steps to prepare the Test System for setup 1 Unpack the test equipment and accessories NOTE This product passed all applicable electrical and mechanical inspections before it wa...

Страница 35: ...powered on and connected before the Agilent 8924C is powered on If the PCS Interface is not powered on first the Agilent 8924C will not detect it 4 Set up the Test Set a Turn on the Test Set b Press...

Страница 36: ...8924 DUPLEX OUT To Agilent 83236 FROM DUPLEX OUT Cable P N 83236 61603 LOCA ADRS RECA SAVE MEAS PRESE INSTRUMENT DATA INCR REF INCR METE INCR AVG LO HI CANC SHIFT k1 k1 k2 k2 k3 k3 k4 ASSIG k5 RELE US...

Страница 37: ...s NOTE This product passed all applicable electrical and mechanical inspections before it was shipped from the factory As soon as you receive it check that it is not damaged and that all accessories a...

Страница 38: ...the rear panel of the PCS Interface as follows GPIB Ser switch GPIB GPIB address 19 for the manual control software and connectivity test 2 Make the front and rear panel connections shown in figure 6...

Страница 39: ...lent 10833D From Agilent 8924 DUPLEX OUT To Agilent 83236 FROM DUPLEX OUT Cable P N 83236 61603 LOCA ADRS RECA SAVE MEAS PRESE INSTRUMENT DATA INCR REF INCR METE INCR AVG LO HI CANC SHIFT k1 k1 k2 k2...

Страница 40: ...setup 1 Unpack the test equipment and accessories NOTE This product passed all applicable electrical and mechanical inspections before it was shipped from the factory As soon as you receive it check...

Страница 41: ...When Used with Agilent 8920B and Controlled via Serial Port for TDMA Subscriber Unit Tests Figure 7 Horizontal stacking configuration side view for the Agilent 8920 Test Configuration Test Set Cellul...

Страница 42: ...done when putting the equipment on the floor the PCS Interface s rear feet must be placed on the same surface as those of the Test Set After placement rotate the handle of the Test Set downwards to p...

Страница 43: ...Cellular Adapter for instructions on connecting it to the Test Set Verify these connections before continuing with these steps 1 Set the GPIB Address Selector on the rear panel of the PCS Interface as...

Страница 44: ...EN BB DATA OUT ANL TRIG OUT BIT CLK OUT SYMBOL CLK OUT OFF CHANNEL PULSE MOD CLK OUT From Agilent 8920 ANT IN To Agilent 83236 TO ANT IN Cable P N 83236 61603 From Agilent 83206 REF IN To Agilent 8920...

Страница 45: ...pare the Test System for setup 1 Unpack the test equipment and accessories NOTE This product passed all applicable electrical and mechanical inspections before it was shipped from the factory As soon...

Страница 46: ...p for example place the Test Set together with the Cellular Adapter on the top of the PCS Interface Position the bumpers feet of the Test Set so that they are stably seated between the forward and rea...

Страница 47: ...utting the equipment on the floor the PCS Interface s rear feet must be placed on the same surface as those of the Test Set After placement rotate the handle of the Test Set downwards to provide suppo...

Страница 48: ...fore continuing with these steps 1 Set the GPIB Address Selector on the rear panel of the PCS Interface as follows GPIB Ser switch GPIB GPIB address 19 for the manual control software and connectivity...

Страница 49: ...CLK OUT OFF CHANNEL PULSE MOD CLK OUT From Agilent 8920 ANT IN To Agilent 83236 TO ANT IN Cable P N 83236 61603 From Agilent 83201 or 83206 REF IN To Agilent 8920 10 MHz REF OUTPUT Supplied with cellu...

Страница 50: ...essories are present see table 2 on page 19 Should this product be damaged or any accessories be missing contact your local Agilent Technologies sales office or the company from which you purchased it...

Страница 51: ...the equipment on the floor the PCS Interface s rear feet must be placed on the same surface as those of the Test Set After placement rotate the handle of the Test Set downwards to provide support for...

Страница 52: ...ng with these steps 1 Set the GPIB Address Selector on the rear panel of the PCS Interface as follows GPIB Ser switch GPIB GPIB address 19 for the manual control software and connectivity test 2 Make...

Страница 53: ...OUT 16 X CHIP CLOCK RF IN OUT TO TEST SET RF IN OUT MAX PWR 60W H 83205A C D M A C E L L U L A R A D A P T E R CDMA TIMEBASE IN EVEN SECOND SYNC IN DATA IN TRIGGER QUALIFIER IN RX TX 19 6608 MHz Clk...

Страница 54: ...essories are present see table 2 on page 19 Should this product be damaged or any accessories be missing contact your local Agilent Technologies sales office or the company from which you purchased it...

Страница 55: ...the equipment on the floor the PCS Interface s rear feet must be placed on the same surface as those of the Test Set After placement rotate the handle of the Test Set downwards to provide support for...

Страница 56: ...n the rear panel of the PCS Interface as follows GPIB Ser switch GPIB GPIB address 19 for the manual control software and connectivity test 2 Make the front and rear panel connections based on your UU...

Страница 57: ...na REF IN External attenuator may be required Consult PCS Interface Specifications POWER FROM DUPLEX OUT TO ANT IN 83236B P C S I N T E R F A C E RF IN OUT RF OUT only TEST SET 1 8 2 0 GHz UUT ON OFF...

Страница 58: ...ator may be required Consult PCS Interface Specifications POWER FROM DUPLEX OUT TO ANT IN 83236B P C S I N T E R F A C E RF IN OUT RF OUT only TEST SET 1 8 2 0 GHz UUT ON OFF User supplied cables From...

Страница 59: ...on EPSON memory card Use the PCMCIA card for the Agilent 8924C or Agilent 8920B use the EPSON memory card for the Agilent 8920A or Agilent 8921A 1 Make proper connections between the Test Set and the...

Страница 60: ...CONTROL I O DIAG OUT GEN BB DATA OUT ANL TRIG OUT BIT CLK OUT SYMBOL CLK OUT OFF CHANNEL PULSE MOD OUT CLK OUT From Agilent 83236 TO ANT IN To Agilent 8920 21 ANT IN Cable P N 83236 61603 From Agilen...

Страница 61: ...X Test screen LOCAL ADRS RECAL SAVE MEAS PRESE INSTRUMENT DATA INCR REF INCR METE INCR AVG LO HI CANC SHIFT k1 k1 k2 k2 k3 k3 k4 ASSIG k5 RELEA USE DAT 7 8 9 4 5 6 1 2 3 0 _ ENTER GHz dB MHz kHz s Hz...

Страница 62: ...elect it The software is now loading Position cursor at Select Procedure Filename and select it Position cursor at Card and select it Press TESTS to display the TESTS Main Menu screen Position cursor...

Страница 63: ...se text area just below the top display line is a two line inverse video area that is used to prompt the user with instructions Main display area below the inverse area is the main display area of the...

Страница 64: ...e Manual Control Soft ware User s Guide for the next steps in making tests using the manual control software If you encountered failures when running the connectivity software If it finds problems the...

Страница 65: ...65 3 P rogramming the PCS Interface using GPIB Control What s included in this chapter Introduction Overview of programming issues PCS Interface programming guide Programming the PCS Interface...

Страница 66: ...formation in this chapter does not apply when an Agilent 8924C is controlling the PCS Interface over the serial port Purpose This chapter is designed to assist Software Application Designers who will...

Страница 67: ...er test UUT port Conversely the PCS Interface takes the UUT s transmitted signal 1710 to 1785 MHz 1805 to 1910 MHz or 1910 to 1990 MHz downconverts it to a frequency the Test Set is capable of analyzi...

Страница 68: ...connection the RF IN OUT port and the RF OUT only port Use the PCS Interface s RF IN OUT port for single port devices for example a subscriber unit or its RF OUT only port for devices with separate RX...

Страница 69: ...analyzer paths In addition the PCS Interface s CPU determines the frequencies required for the Test Set s RF generator and analyzer These values are returned over GPIB and are then used by the contro...

Страница 70: ...and computes a level which should be set at the Test Set s DUPLEX OUT port This level is returned over GPIB and is then used by the controlling software to set the Test Set s RF generator level The au...

Страница 71: ...mputes and sets the analyzer path attenuator accordingly NOTE This is a one time setting and does not place the PCS Interface in an auto ranging mode If the TX signal level changes the attenuation set...

Страница 72: ...later in this chapter Power measurements After the UUT s transmitter signal is applied to the RF IN OUT port and the desired trigger and data collection conditions are set in the PCS Interface the po...

Страница 73: ...moves the dc drift of the power detector to provide accurate power measurements of signal levels less than 0 dBm During this procedure the UUT connected to the PCS Interface s RF IN OUT port is discon...

Страница 74: ...Set s DUPLEX OUT port Before using the generator for receiver testing the need for a new generator level temperature compensation should be checked The controlling software queries the PCS Interface...

Страница 75: ...d Examples given in this chapter will include the instrument to be programmed the command string and the reference page in this manual for the given command string For example To the PCS Interface COM...

Страница 76: ...ation set up and measurement steps required in control software when using a PCS Interface in a test system Figure 24 Start Initialize the Set Up and More End System Measure Y N Place the Test Set and...

Страница 77: ...t instrument settings 1 Use the following command string to reset the Test Set and query for completion To the Test Set RST OPC 2 Use the following command string to reset the PCS Interface and query...

Страница 78: ...ANT IN ports Placing the Test Set in Manual Tune mode will provide accurate frequency error results when transmitter frequency measurements are made 1 Use the following command string to set the Test...

Страница 79: ...stem Figure 26 Set Up and Measure Start Initialize the Set Up and More End System Measure Y N Initialize Set Up the UUT Ports Set the Test Frequencies and Levels for RX or TX testing Test Procedure Fl...

Страница 80: ...the first time requested following the RST command Set Up the UUT ports The internal signal path switches of the PCS Interface can be controlled to route the Test Set s connections to or from the UUT...

Страница 81: ...the Test Set s DUPLEX OUT frequency To the Test Set RFG FREQ Duplex_freq NOTE Table 14 on page 138 describes the frequency conversions automatically made by the PCS Interface Set the generator level...

Страница 82: ...uency to be expected at the Test Set s ANT IN port To the PCS Interface TX TSET FREQ test frequency page 120 2 Define a local constant for the frequency obtained in step 1 For example Ant_freq 3 Use t...

Страница 83: ...r receiver testing Since the Test System s source is used for receiver testing the generator level temperature compensation should be checked and performed if necessary prior to making a receiver meas...

Страница 84: ...UPLEX OUT port To the PCS Interface COMP TEMP LEV page 103 e Define a local constant for the level obtained in step d For example Duplex_lev f Use the following command string to set the Test Set s ge...

Страница 85: ...IN port and to query for completion To the PCS Interface TX OUTP LEV ADJ OPC page 119 NOTE Do not have a signal generator level greater than 50 dBm at the RF IN OUT port when adjusting the attenuator...

Страница 86: ...e instrument is ready after being instructed to make a power measurement This type should be used for any continuous signal measurement Amplitude trigger Measurements start when the PCS Interface dete...

Страница 87: ...r conditions have been satisfied To the PCS Interface TX INP POW PDET SAMP IGN desired num ber page 117 b Use the following command string to define the number of samples to collect ranging from 1 to...

Страница 88: ...plies 3 29 samples symbol The values for IGNORE and LENGTH provide power measurements compliant with the IS 137 specification of at least 140 symbols between symbols 6 and 162 in the burst Table 9 Exa...

Страница 89: ...l from the Cellular Adapter is used for the external trigger and the subscriber unit is locked to the Test System s frame clock transmitter power measurements can be made on any of the 6 TDMA slots by...

Страница 90: ...command string is used to make a power measurement When performing the power measurement the RF generator level should be set to less than 50 dBm if the RF IN OUT port is used or set to less than 30...

Страница 91: ...91 4 G PIB Commands What s included in this chapter Introduction The Syntax of Program Messages Reference Information for each GPIB command GPIB Command Summary GPIB Commands...

Страница 92: ...mand tree is called the root command or simply the root This is similar to the terminology used for the directory path of DOS files To indicate the command at the beginning of the tree the path from t...

Страница 93: ...reviations for commands when describing them However the letters that are used in the abbreviation of the command appear in uppercase and the letters that must be omitted to form the abbreviation appe...

Страница 94: ...Program Message Terminator The program message terminators are newline END and newline END END is used to indicate the end of a program message END means that an EOI End Of Identify was transmitted s...

Страница 95: ...ter string as shown in the example below IDN Commas and semicolons are used to separate data in response messages Commas are used to separate multiple return values in response to a Query command Semi...

Страница 96: ...tial expression with a space after the e 7 89E 01 exponential expression with an E or an e 5 only to the right of the decimal point There is a limit to the numeric formats that can be used when using...

Страница 97: ...pecified in numeric Query only 3 numeric 130 to 10 dBm for RF PATH 1 130 to 20 dBm for RF PATH 2 4 Response to the Query numeric dBm new line END 1 Command name and required parameter When a parameter...

Страница 98: ...xxxxxxxx REV yy yy newline END where xxxxxxxxxx is a serial number and yy yy is a firmware version OPT Returns the options that are installed Query only NO OPTION The frequency range is limited See Ch...

Страница 99: ...MP REQ RES 1 COMP TEMP REQ RES 1 RF PATH 0 RX OUTP ATT 70 dB RX OUTP FREQ 1930 MHz RX OUTP LEV 130 dBm TX INP FREQ 1850 MHz TX INP POW PDET SAMP AVER 1 for the Immediate Trigger 5 for the Amplitude or...

Страница 100: ...e it to the value stored the last time dc offset values were measured When the difference is greater than or equal to the value set by the COMP PDET DCOF REQ RES command the PCS Interface stores the c...

Страница 101: ...than or equal to 0 in 0 1 C steps If 0 is set the temperature compensation is always required Response to the Query numeric new line END COMPensation RFG1 TEMPerature EXECute This command performs the...

Страница 102: ...quency 1750 MHz 3 1805 to 1850 MHz band Compensation frequency 1830 MHz 4 1850 to 1910 MHz band Compensation frequency 1880 MHz 5 1930 to 1990 MHz band Compensation frequency 1960 MHz numeric 800 to 8...

Страница 103: ...e the COMP TEMP EXEC command to execute the RF generator temperature compensation Response to the Query numeric dBm new line END COMPensation RFG1 TEMPerature PDETector SAMPle AVERage FACTor numeric S...

Страница 104: ...the threshold value of the temperature change to judge whether an update of the RF generator compensation factors are required This value is used by the COMP TEMP REQ STAT command numeric Greater tha...

Страница 105: ...z band Compensation frequency 1750 MHz 3 1805 to 1850 MHz band Compensation frequency 1830 MHz 4 1850 to 1910 MHz band Compensation frequency 1880 MHz 5 1930 to 1990 MHz band Compensation frequency 19...

Страница 106: ...IN OUT 1710 to 1990 MHz to UUT RF OUT only 1710 to 1990 MHz to UUT EXT TRIG IN Frame Clk from GPIB to Agilent 892NX Serial to Agilent 8924C DET OUT to Agilent 8924C REF IN from Agilent 892NX REF OUT...

Страница 107: ...tenuator is set to 40 dB and the RX attenuator is set to 70 dB Response to the Query 1 0 1 2 newline END 1 indicates that it is an undefined signal path Table 11 RF Path Setting Parameter SW 1 SW 2 RF...

Страница 108: ...90 or 990 MHz to 1110 MHz 10 MHz step 0 to 70 dB 10 dB step RF IN OUT 1710 to 1990 MHz to UUT RF OUT only 1710 to 1990 MHz to UUT EXT TRIG IN Frame Clk from GPIB to Agilent 892NX Serial to Agilent 892...

Страница 109: ...t LEVel Returns the level required at the FROM DUPLEX OUT port Query only Before executing this command specify the level of the signal which you want to output at the RF IN OUT or RF OUT only port us...

Страница 110: ...UTP LEV command Response to the Query AUTO HOLD new line END RX RFG1 OUTPut FREQuency numeric HZ KHZ MAHZ MHZ GHZ Specifies the frequency of the signal which you want to output from the RF IN OUT or R...

Страница 111: ...the functions of both the RX OUTP LEV and RX INP LEV command RX RFG1 TSET FREQuency numeric HZ KHZ MAHZ MHZ GHZ Specifies the frequency of the signal which you want to output from the RF IN OUT or RF...

Страница 112: ...rt and returns the level required at the FROM DUPLEX OUT port when the level of the wanted signal at the RF IN OUT or RF OUT only is specified in numeric Query only numeric 130 to 10 dBm for RF PATH 1...

Страница 113: ...PRES Command Settings Command Setting COMP PDET DCOF REQ RES Current Setting COMP PDET TEMP REQ RES Current Setting COMP TEMP REQ RES Current Setting RF PATH 0 RX OUTP ATT 70 dB RX OUTP FREQ 1930 MHz...

Страница 114: ...1110 MHz 10 MHz step 0 to 70 dB 10 dB step RF IN OUT 1710 to 1990 MHz to UUT RF OUT only 1710 to 1990 MHz to UUT EXT TRIG IN Frame Clk from GPIB to Agilent 892NX Serial to Agilent 8924C DET OUT to Agi...

Страница 115: ...perature compensation for the power detector measures the average power of the burst of the signal input to the RF IN OUT port and returns the value to the controller The trigger source which decides...

Страница 116: ...y only Response to the Query numeric dBm or W new line END The unit suffix dBm or W is specified by the TX INP POW UNIT command The default is dBm If the measured data is invalid 130 dBm or 0 W is ret...

Страница 117: ...sponse to the Query numeric new line END NOTE The following equation shows the relationship between the IGNORE value set by the TX INP POW PDET SAMP IGN command and the LENGTH value set by the TX INP...

Страница 118: ...ely immediate trigger Response to the Query AMPTD EXT IMM new line END TX INPut POWer UNIT DBM W Specifies the measurement units for the input signal s power result returned by the TX INP POW command...

Страница 119: ...zer LEVel numeric DBM Specifies the peak level which you want to output to the TO ANT IN port Use the TX OUTP LEV ADJ command to automatically adjust the TX attenuator to output this signal level from...

Страница 120: ...s command is valid only when the setting of RF PATH is 1 or 2 numeric 800 to 960 MHz or 1710 to 1990 MHz The default unit is Hz Response to the Query numeric dB new line END TX TSET FREQuency numeric...

Страница 121: ...on TEMPerature REQuest RESolution RFG1 TEMPerature EXECute FREQuency LEVel PDETector SAMPle AVERage FACTor LENGth REQuest RESolution STATe 0 1 2 3 4 5 numeric numeric 0 0 1 2 3 4 5 numeric Query only...

Страница 122: ...Query only No Query TX INPut FREQuency POWer POWer EXECute EXECute PDETector SAMPle AVERage FACTor IGNore LENGth TRIGger SOURce UNIT OUTPut RFANalyzer ATTenuator FREQuency LEVel ADJust PATH ILoss TSE...

Страница 123: ...123 5 S pecifications Specifications...

Страница 124: ...he temperature compensation has been implemented and are valid over the entire operating range unless otherwise noted Supplemental Characteristics are intended to provide additional information useful...

Страница 125: ...E To meet generator output path specifications the input signal must be from the Agilent 8920A B Agilent 8921A D or Agilent 8924C Test Set with the following characteristics Frequency Range 800 MHz to...

Страница 126: ...to 960 MHz Conversion Path 1710 MHz to 1785 MHz 1805 MHz to 1910 MHz 1930 MHz to 1990 MHz Frequency Settling Time 10 ms Output RF IN OUT Port 130 to 20 dBm RF OUT only Port 130 to 10 dBm Output Level...

Страница 127: ...ectral Purity Offset 5 kHz For carrier levels 100 dBm SSB Phase Noise 100 dBc Hz at 20 kHz offset from carrier Table 13 Spurious Type of spurious 800 to 960 MHz 960 to 1710 MHz 1710 to 1990 MHz Harmon...

Страница 128: ...128 Chapter 5 Specifications Analyzer Input Path Analyzer Input Path RF IN OUT connector Figure 29 Block Diagram Analyzer Input Path Agilent 83236B Agilent 892NX...

Страница 129: ...MHz Input Max Input Level Subscriber Unit Test Single carrier TDMA and FM 40 dBm 10 W CDMA 37 dBm 5 W Base Station Test Single carrier TDMA and FM 40 dBm 10 W CDMA and multi carrier 30 dBm 1 W Max Pe...

Страница 130: ...ment Range Subscriber Unit Test Single carrier TDMA and FM 13 dBm to 40 dBm 50 W to 10 W CDMA 13 dBm to 37 dBm 50 W to 5 W Base Station Test Single carrier TDMA and FM 13 dBm to 40 dBm 50 W to 10 W CD...

Страница 131: ...evel Range 5 dBm to 10 dBm REF OUT Output Frequency 10 MHz Output Level 1 dBm Accuracy Buffered signal from REF IN or 10 MHz 10 ppm if no external reference is connected to REF IN Remote Control GPIB...

Страница 132: ...5 W Base Station Test Single carrier TDMA and FM 40 dBm 10 W CDMA and multi carrier 30 dBm 1 W Max Peak Instantaneous Signal 30 V RF OUT only Output to UUT Output Impedance 50 nominal Connector Type...

Страница 133: ...vel Range 70 dBm to 7 dBm Input Impedance 50 nominal Connector Type Type N F SWR 1 3 1 REF IN Input from the Test Set Input Frequency 10 MHz Input Level Range 5 dBm to 10 dBm Input Impedance 50 nomina...

Страница 134: ...TDMA External Trigger Signal Required for TDMA RF input level range 13 dBm to 5 dBm Input Level TTL Input Impedance 1 k at 1 MHz nominal Triggering Type Positive Edge Connector Type SMC M DET OUT Out...

Страница 135: ...mm Weight 5 6 kg Operating Temperature 0 C to 55 C Operating Humidity 15 to 95 RH 40 C Operating Altitude 0 to 2000 meters Non operating Temperature 40 C to 70 C Non operating Humidity to 90 RH 65 C...

Страница 136: ...4 3 1995 EN 61000 4 3 1996 3 V m 80 AM 27 1000 MHz IEC 61000 4 4 1995 EN 61000 4 4 1995 0 5 kV signal lines 1 kV power lines IEC 61000 4 5 1995 EN 61000 4 5 1995 0 5 kV line line 1 kV line ground IEC...

Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...

Страница 138: ...810 to 820 MHz 1820 to 1830 MHz 1010 MHz 810 to 820 MHz 1830 to 1840 MHz 1020 MHz 810 to 820 MHz 1840 to 1850 MHz 1030 MHz 810 to 820 MHz 1850 to 1860 MHz 1020 MHz 830 to 840 MHz 1860 to 1870 MHz 1030...

Страница 139: ...ion Table 15 Frequency Conversion Table for Signal Analyzer Input Frequency from UUT MHz LO Setting MHz Required Input Frequency to Test Set s ANT IN port MHz 800 to 960 MHz bypass path 800 to 960 MHz...

Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...

Страница 141: ...141 B M anual Changes Appendix B Manual Changes...

Страница 142: ...g the information in this manual to correspond to the PCS Interface units that were manufactured before this manual was printed If the serial number of your PCS Interface matches one of the serial num...

Страница 143: ...this manual or those noted in table 16 and table 17 write down the changes in your operation manual corresponding to the yellow revision notes In addition to this information the yellow revision notes...

Страница 144: ...s Serial Numbers The serial number of your PCS Interface is marked on the serial number plate such as the one shown in figure 30 The plate is attached to the rear panel Figure 30 Serial Number Plate A...

Страница 145: ...n frequency 880 MHz 2 1710 to 1785 MHz band Compensation frequency 1750 MHz 3 1805 to 1850 MHz band Compensation frequency 1830 MHz 4 1850 to 1910 MHz band Compensation frequency 1880 MHz 5 1930 to 19...

Страница 146: ...Compensation frequency 1750 MHz 3 1805 to 1850 MHz band Compensation frequency 1830 MHz 4 1850 to 1910 MHz band Compensation frequency 1880 MHz 5 1930 to 1990 MHz band Compensation frequency 1960 MHz...

Страница 147: ...se the RX TSET FREQ command which has the functions of both the RX OUTP FREQ and RX INP FREQ commands RX RFG1 TSET FREQuency numeric HZ KHZ MAHZ MHZ GHZ Specifies the frequency of the signal which you...

Страница 148: ...value of the insertion loss of the signal path for the RF signal analyzer when the frequency of the input signal to the RF IN OUT port is specified by numeric This command is used to calculate the RF...

Страница 149: ...following characteristics Frequency Range 810 MHz to 995 MHz Level Range 70 dBm to 7 dBm Frequency Frequency Range Through Path 824 MHz to 849 MHz 869 MHz to 894 MHz Conversion Path 1710 MHz to 1785...

Страница 150: ...Frequency Frequency Range Through Path 824 MHz to 849 MHz 869 MHz to 894 MHz Conversion Path 1710 MHz to 1785 MHz 1805 MHz to 1910 MHz 1930 MHz to 1990 MHz Power Measurement Frequency Range 824 MHz to...

Страница 151: ...30 MHz 1010 MHz 810 to 820 MHz 1830 to 1840 MHz 1020 MHz 810 to 820 MHz 1840 to 1850 MHz 1030 MHz 810 to 820 MHz 1850 to 1860 MHz 1020 MHz 830 to 840 MHz 1860 to 1870 MHz 1030 MHz 830 to 840 MHz 1870...

Страница 152: ...onversion Table for Signal Analyzer Input Frequency from UUT MHz LO Setting MHz Required Input Frequency to Test Set s ANT IN port MHz 824 to 894 MHz bypass path 824 to 894 MHz 1710 to 1758 MHz 1060 M...

Страница 153: ...the PCS Interface These error messages are output through the GPIB interface A description of its contents and how to intervene to correct the error is provided for each error message When an error is...

Страница 154: ...ds were out of range 102 Current path is invalid for the command The command that was sent could not be executed due to the current RF PATH specification 100 Command error A general syntax error that...

Страница 155: ...to 119 is generated 111 Header separator error The header separator was incorrect This message is displayed if a character other than those allowed in headers immediately follows a header This messag...

Страница 156: ...large An exponent was too large The largest allowable exponent is 32000 See IEEE 488 2 7 7 2 4 1 124 Too many digits Numeric data was too long The maximum allowable length is 255 digits 128 Numeric d...

Страница 157: ...ld not be recognized was sent 144 Character data too long Character data was too long The maximum allowable length is 12 characters 148 Character data not allowed Character data may not be used 150 St...

Страница 158: ...be executed due to the current settings of the device See IEEE 488 2 6 4 5 3 and 11 5 1 1 5 222 Data out of range A numeric parameter of an GPIB command is out of range 224 Illegal parameter value The...

Страница 159: ...is missing This error may be generated if an option is not installed 310 System error A system error was generated 311 Memory error A memory error An error occurred in the memory of this device 330 Se...

Страница 160: ...error A Query command was invalid 410 Query INTERRUPTED A DAB or a GET command was received before the response to a Query command was completed 420 Query UNTERMINATED An incomplete program message wa...

Страница 161: ...ks the basic functionality of each instrument in the Test System frame A basic timing interval For CDMA access paging and traffic channels a frame is 20 ms long for the sync channel a frame is 26 666...

Страница 162: ...ond the USER keys on the Test Set s front panel The softkey s function can be activated by pressing its corresponding USER key subscriber unit See mobile station terminal See port TDMA Time Division M...

Страница 163: ...llation 31 L local oscillator definition 161 M manual changes 141 mobile station See subscriber unit N nominal 124 numeric 96 O options 22 over the air definition 161 P PCMCIA media definition 162 PCS...

Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...

Отзывы: