145
Appendix B, Manual Changes
Change 1
Change 1
Change the following parts if the response to the query of the “*OPT” command
is “NO OPTION”.
"COMPensation Subsystem" on page 100
COMPensation[:RFG1]:TEMPerature:FREQuency?
ëû
[{0|1|2|3|4|5|
<numeric>
[HZ|KHZ|MAHZ|MHZ|GHZ]}]
This command returns the frequency required at the FROM DUPLEX OUT port
for the temperature compensation for the RF signal generator to provide the
optimum output level accuracy. If the frequency band or value is not specified,
the frequency value specified by the
RX:OUTP:FREQ
is used. (Query only)
0:
824 to 849
MHz band (Compensation frequency: 840 MHz)
1:
869 to 894
MHz band (Compensation frequency: 880 MHz)
2:
1710 to 1785 MHz band (Compensation frequency: 1750 MHz)
3:
1805 to <1850 MHz band (Compensation frequency: 1830 MHz)
4:
1850 to 1910 MHz band (Compensation frequency: 1880 MHz)
5:
1930 to 1990 MHz band (Compensation frequency: 1960 MHz)
<numeric>:
824 to 849
MHz,
869 to 894
MHz, 1710 to 1785 MHz, 1805 to
1850 MHz, 1850 to 1910 MHz, 1930 to 1990 MHz (Compensation
frequency is rounded off to 840 MHz, 880 MHz, 1750 MHz,
1830 MHz, 1880 MHz, or 1960 MHz.) If the entered value is out of
range, the value is rounded off to the appropriate limit value. (The
default unit is “Hz”.)
Supply the signal whose frequency and level are returned by this command and
the
COMP:TEMP:LEV?
command to the FROM DUPLEX OUT port. Then
execute the
COMP:TEMP:EXEC?
command to execute the RF generator
temperature compensation.
•
Response to the Query
{numeric (Hz)} <new line><^END>
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...