111
Chapter 4, GPIB Commands
RX Subsystem
RX[:RFG1]:OUTPut:LEVel
ëû
<
numeric
>[DBM]
Specifies the level of the signal which you want to output from the RF IN/OUT or
RF OUT only port.
<numeric>
−
130 to
−
10 dBm for RF:PATH 1
−
130 to
−
20 dBm for RF:PATH 0 and 2
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric (dBm)} <new line><^END>
NOTE:
Use the
RX:TSET:LEV?
command which has the functions of both the
RX:OUTP:LEV
and
RX:INP:LEV?
command.
RX[:RFG1]:TSET:FREQuency?
ëû
<
numeric
>[HZ|KHZ|MAHZ|MHZ|GHZ]
Specifies the frequency of the signal which you want to output from the RF IN/
OUT or RF OUT only port, and returns the frequency required at the FROM
DUPLEX OUT port, when the frequency of the wanted signal at the RF IN/OUT
or RF OUT only is specified in
<
numeric
>
. (Query only)
<numeric>:
800 to 960 MHz, 1710 to 1785 MHz, 1810 to 1910 MHz, or
1930 to 1990 MHz
(The default unit is “Hz”.)
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric (Hz)} <new line><^END>
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...