Cleaning Information
Which Cleaning Procedure should I use ?
98
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Which Cleaning Procedure
should I use ?
Light dirt
If you just want to clean away light dirt, observe the following
procedure for all devices:
• Use compressed air to blow away large particles.
• Clean the device with a dry cotton swab.
• Use compressed air to blow away any remaining filament left by the
swab.
Heavy dirt
If the above procedure is not enough to clean your instrument, follow
one of the procedures below. Please consult
for this Instrument” on page 90
for the procedure relevant for this
instrument.
If you are unsure of how sensitive your device is to cleaning, please
contact the manufacturer or your sales distributor
How to clean connectors
Cleaning connectors is difficult as the core diameter of a single-mode
fiber is only about 9
µ
m. This generally means you cannot see streaks
or scratches on the surface. To be certain of the condition of the
surface of your connector and to check it after cleaning, you need a
microscope.
In the case of scratches, or of dust that has been burnt onto the surface
of the connector, you may have no option but to polish the connector.
This depends on the degree of dirtiness, or the depth of the scratches.
This is a difficult procedure and should only be performed by a skilled
person, and as a last resort as it wears out your connector.
W A R N I N G
Never look into the end of an optical cable that is connected to an
active source.
Содержание 81610A
Страница 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Страница 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Страница 51: ...Accessories Modules and Options 48 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Страница 57: ...Specifications Definition of Terms 54 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Страница 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Страница 113: ...Cleaning Information Other Cleaning Hints 110 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 115: ...Cleaning Information Other Cleaning Hints 112 Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition ...
Страница 116: ......
Страница 117: ... Agilent Technologies GmbH 2000 2002 Printed in Germany February 20 2002 81610 90A15 81610 90A15 www agilent com ...