Installation
8-25
a. Connect the gray color-coded connector to the wall box receptacle.
Align the slots in the connector with the keys in the receptacle and
push in as far as possible. Secure the connector to the receptacle by
tightening the retaining ring onto the receptacle.
b. Connect the brown color-coded connector on the other end of the LDC
to the mating brown color-coded connector located on the instrument
(bedside monitor). Align the slots in the connector with the keys in the
receptacle and push in as far as possible. Secure the connector to the
receptacle by tightening the retaining ring onto the receptacle.
c. Connect any additional patient related instruments in series using
other local distribution cables. Observe the upstream and downstream
color-coding scheme as described above.
Caution
The sum total of all the lengths of LDCs must be less than 15.2 meters
(50 ft.).
Terminate Variable Length LDC
The following procedures can be used to terminate the variable length
LDC cable kit Agilent Technologies part number 78599AI-J50 and, also, to
terminate the 2-connector kit, Agilent Technologies part number 785-99Al-
J52 and 78599Al-J54.
1.
Measure and cut the open end of the LDC to the desired length. Leave
at least 0.3 meters (1 ft.) overrun to facilitate termination and
instrument servicing.
2.
Insert the open end of the cable through the connector hardware as
illustrated below.
3.
Strip off approximately 2.5 cm (1 in.) of insulation from the cable end
exposing the braided shield.
4.
Use care not to cut or damage the braided shield.
5.
Unbraid and peel back the braided shield. Separate the drain/ground
wires from the shield and twist together. Fold back the braided shield
over the cable insulation.
6.
Cut off the foil and filler strips from the cable end.
7.
Strip off approximately 6.3 mm (1/4 in.) of insulation from each of the
four wire ends to allow the bared end to be soldered onto the
connector contacts.
8.
Solder the four wire ends and the drain wire to the respective terminal
contacts as indicated below.
Содержание 78581B
Страница 46: ...ACC Maintenance 4 2 ...
Страница 54: ...Installation Responsibilities Customer and Agilent Technologies 6 4 ...
Страница 90: ...Procedures 8 28 ...
Страница 98: ...Contacts 10 6 ...