2-13
Application Tutorials
Tutorial 4: Measure Laser Turn-on Delay
Tutorial 4: Measure Laser Turn-on Delay
On-screen markers can be used to measure both amplitude and time separa-
tion in eye-diagrams. The compatible modes for markers are eye, eyeline, and
pattern.
Change to Eyeline Mode
1
Change to eyeline mode, then turn filtering on by pressing:
INSTR PRESET
,
Setup
,
diagram
,
EYELINE
,
eyefltr ON OFF
ON,
BIT INTVL
and enter
1
2
Set the persistence to infinite by pressing:
Traces
,
persist
,
INFINIT
,
Trg,Cal
3
After several traces have been displayed, press:
CONT STOP
STOP
Turn on the markers
4
Press
X1
and
X2
to activate markers 1 and 2.
5
Turn the front-panel knob to move the markers to different transition crossing
points on the waveform as show in the following figure.
This provides an easy method to check the peak-to-peak difference in the laser
turn-on time measured at the crossing point. Notice in the following figure, a
delta reading of 30 ps is displayed at the top of the screen.
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...