Application Tutorials
Tutorial 2: Measure in Optical Power Units
Tutorial 2: Measure in Optical Power Units
The analyzer display has the ability to show optical units referenced to the
input of the optical-to-electrical converter. This changes the channel and
marker readouts to watts/div.
1
Set the analyzer to a known state by pressing:
INSTR PRESET,
Trg,Cal,
CH1 EXT ATTEN
responsivity value
,
V/Watt
For example, the following figure shows 350 V/Watt entered. Notice the soft-
key label changes to CH1 RSPVTY (responsivity).
2
To use the markers to read optical power, press:
Markers
,
Y1 (--)
3
Adjust the marker line to the peak of the response. For example, the following
figure shows a peak optical power of 1.5 mW.
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...