1-25
Getting Started
Steps for Setting Up Eye-Diagram Analysis
70841A/B appears.
c
Press
ADJUST ROW
, and rotate the knob to move the box to the row where
the 70841A/B appears.
d
Press
ASSIGN BOTH
.
4
Use the displayed softkey menus to set the pattern generator to the desired
settings. For an example, refer to the paragraph "Configure the data signal" in
Chapter 2, “Application Tutorials”
5
To return control to the eye-diagram analyzer, press:
DISPLAY,
NEXT INSTR
6
To return to the eye-diagram analyzer personality menus, press:
USER
To Manually Set
the Trigger Level
If a pattern generator other than the 70841A/B is used, you must use the fol-
lowing procedure to manually set the trigger level.
1
Select trace two and turn the display on by pressing:
>
prev menu,
select:
,
TR2
,
display ON/OFF
ON
2
Automatically scale the trigger trace and set the trigger level by pressing:
AUTO-SCALE
,
Trg,Cal
,
LEVEL
Use the knob to position the trigger level indicator line so it crosses an edge.
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...