3-63
Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
Matching Delay Between Channels
Significant delays can occur between the signals on channel 1 and channel 2
when the electrical lengths of external cabling varies. This results in a phase
offset between the two channels. The microwave transition analyzer module
offers the capability to add delay to either channel 1 or channel 2.
The following figure shows added delay on channel 2.
Use the AUTO SKEW softkey to automatically compensate the delay on chan-
nel 2. Automatic skewing assumes two equal phase signals input to channels 1
and 2. If measurements are to be performed at multiple frequencies, it is rec-
ommended that the AUTO SKEW function be performed first at a low fre-
quency and then at higher frequencies to ensure better resolution.
Delay can be manually compensated using the CHAN2 SKEW softkey. For
convenience, the SEC/DIV softkey is provided so that the time scale can be
changed.
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...