Eye-Diagram Analyzer Reference
Saving to Mass Storage
6
Issue the following Instrument BASIC command to copy the files from the RAM
disk to the new card:
COPY "*:MEMORY,0,0" TO ":EXTERNAL,904"
To Catalog all
Files
Use this procedure to catalog default memory. To select default memory, refer
to
“Selecting and Formatting Memory” on page 3-34
1
If you plan to catalog the files on a memory card, insert the card in the front-
panel card slot.
2
Display a catalog of the files by pressing:
MENU
,
page 1 of 2
,
States
,
more 1 of 2,
mass storage
Remember to insert the card before cataloging a memory card. If the message
"
6218 catalog open failed
" is displayed, the card is missing. The HP-
MSIB address for the card is the same address as the display’s GPIB address
and is normally set to 4. If the address is not correct, enter the correct address
using the numeric keypad.
To Save a File
Use this procedure to save a file in the default memory. To select default mem-
ory, refer to
“Selecting and Formatting Memory” on page 3-34
1
Display the Mass Storage menu by pressing:
MENU
,
page 1 of 2
,
States
,
more 1 of 2,
mass storage
,
save
2
Perform one of the following:
• If you are saving an instrument state, user-correction, or histogram, press the
appropriate softkey.
• If you are saving a trace, press
save trace
. Select the trace and then use the
VEC|FMT
softkey to determine if the trace data is saved as complex data
(VEC) or scalar data (FMT). Press save.
• If you are saving a channel calibration, mask, limit line, user-defined key, all
recall registers, press
save misc
and then the appropriate softkey.
3
Enter the desired file name.
4
To save the file, press:
ENTER LINE
An HP-IL keyboard is recommended to simplify the task of entering alphanu-
meric information. This keyboard plugs into the jack provided on the display’s
front panel. Order the Agilent 46021A keyboard.
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...